Micro Solutions

Analytical solutions for your laboratory

RU
Molecular diagnostic system NeXgen 16HP

Zinext

Molecular diagnostic system NeXgen 16HP

Compact system “from test to result”: extraction and multiplex qPCR detection in one cycle

NeXgen 16HP is a compact Zinexts test-to-result system: automatic sample preparation and multiplex qPCR detection for 1–16 samples per launch.

Request a quotation

The device homogenizes the sample using proprietary tip spin technology and releases the nucleic acids with magnetic particles inside the same tip. The built-in module then automatically prepares the PCR reaction in the cooled unit and seals the tubes with a proprietary mechanism, after which the 32-well thermocycler conducts a multiplexing. qPCR detection through 5 or more channels.

The system processes 1 to 16 samples per launch, including complex samples up to 10 ml (urine, saliva, plasma) thanks to built-in fluid level determination. The 32-hole thermocycle supports separate (twin) qPCR reactions and is compatible with both lyophilized and liquid reagents. integration with LIS and is open to third-party analysis formats.

Complete cycle from test to result

  • Patented tip spin technology homogenizes the sample right at the tip
  • Extraction of nucleic acids by magnetic particles inside the tip
  • Built-in module of preparation of PCR reaction with cooled block for 14 tubes and automatic patented sealing of tubes

PCR detection

  • 32-well thermocycler with the possibility of dividing the sample for double qPCR reactions
  • High-multiplex qPCR – 5 or more detection channels
  • Compatible with lyophilized and liquid reagents qPCR

Flexibility and integration

  • Determination of the fluid level when working with complex samples up to 10 ml (urine, saliva, plasma)
  • Full bidirectional integration with LIS
  • Open system compatible with different analysis formats

Throughput

ParameterNeXgen 16HP
Number of samples per launch1–16
Maximum sample sizeup to 10 ml (urine, saliva, plasma and other complex samples)

Extraction and detection

ParameterNeXgen 16HP
Nucleic acid extractionmagnetic particles at the tip (in-tip), spin-homogenization
PCR modulecooled unit for 14 tubes, automatic closure
Thermocycler32-well, support for split (twin) qPCR reactions
Multiplexity of detection5 or more channels
Compatible reagentsLiquid and qPCR reagents

Integration

ParameterNeXgen 16HP
Liaison to LISbidirectional
Platform opennessCompatible with third-party analysis formats