Gas chromatograph YL6500 GC

Flexible Gas Chromatography: Up to three detectors and flow division up to 7,500:1

Gas chromatograph YL6500 GC

Product overview

Flexible Gas Chromatography: Up to three detectors and flow division up to 7,500:1

with a thermostat of columns up to 450 °C and electronic control of gases APC

YL6500 GC is a Young In Chromass gas chromatograph with a flexible configuration of injectors and detectors, used including for the determination of ethanol and higher alcohols by vapor-phase method.

The system supports split/splitless and on-column sample input, as well as up to three detectors simultaneously — PID, catarometer, thermoionic, flame-photometric, pulsed discharge detector or micro-electronic capture detector. Electronic gas control APC controls up to 18 channels, and data is processed in the program. YL-Clarity with 21 CFR Part 11 support

The thermostat of the columns operates from 4 °C above ambient temperature to 450 °C and above, with optional cryogenic cooling to −80 °C (liquid nitrogen) or −40 °C (CO2); stability ±0.01 °C in isothermal mode. Capillary injector provides flow division up to 7,500:1 for columns with a diameter of 0.05–0.53 mm. Detection limits: PID - 1.5 pg C/s, micro-ECD - 4.4 fg/s.

Key figures

3

detector

7 500:1

max.

450 °C

max. oven temperature

18

channel

Applications

Models and configurations

Sampling

Detectors.

Column thermostat and gas control

Sampling (injectors)

Injector, max. temperature450 °C
Nozzle injector, pressure0,001–100 psi
Nozzle injector, flow0.01–100 ml/min
Capillary Injector (split/splitless), Maximum Temperature450 °C
Flow divisionup to 7,500:1
Diameter of capillary columns0.05–0.53 mm

Column thermostat

Temperature range4 °C above the ambient to ≥450 °C
Cryogenic cooling (option)up to -80 °C (liquid nitrogen) / up to -40 °C (CO2)
Temperature stability±0.01 °C isothermically; ±0.1 °C in gradient

Detectors.

PID (FID)up to 450 °C, detection limit of 1.5 pg C/s, dynamic range 107
Catarometer (TCD)up to 400 °C, detection limit of 2.5 ng/ml
Thermoion Detector (NPD)up to 400 °C, dynamic range >104
Flame and Photometric Detector (FPD)up to 300 °C (limit 250 °C), sensitive to sulfur and phosphorus
Pulse discharge detectorup to 400 °C, helium ionization or electron capture modes
Micro-detector of electronic capture (μECD)up to 450 °C, detection limit 4.4 fg/s

Gas control and SO

Electronic Gas Control (APC)up to 6 blocks, 18 channels; N2, He, H2, air, Ar/CH4
SoftwareYL-Clarity, conforming to 21 CFR Part 11, up to 4 chromatographs of 12 detectors in the network