Young In Chromass
Gas chromatograph YL6500 GC
Flexible Gas Chromatography: Up to three detectors and flow division up to 7,500:1
YL6500 GC is a Young In Chromass gas chromatograph with a flexible configuration of injectors and detectors, used including for the determination of ethanol and higher alcohols by vapor-phase method.
Request a quotationThe system supports split/splitless and on-column sample input, as well as up to three detectors simultaneously — PID, catarometer, thermoionic, flame-photometric, pulsed discharge detector or micro-electronic capture detector. Electronic gas control APC controls up to 18 channels, and data is processed in the program. YL-Clarity with 21 CFR Part 11 support
The thermostat of the columns operates from 4 °C above ambient temperature to 450 °C and above, with optional cryogenic cooling to −80 °C (liquid nitrogen) or −40 °C (CO2); stability ±0.01 °C in isothermal mode. Capillary injector provides flow division up to 7,500:1 for columns with a diameter of 0.05–0.53 mm. Detection limits: PID - 1.5 pg C/s, micro-ECD - 4.4 fg/s.
Sampling
- Nozzle injector up to 450 °C, pressure 0.001-100 psi, flow 0.01-100 ml/min
- Capillary injector (split/splitless) up to 450 °C, flow division up to 7,500:1, columns 0.05–0.53 mm
- On-column injector with temperature programming up to 5 steps
Detectors.
- Up to three detectors simultaneously: PID (FID), catarometer (TCD), thermoionic (NPD), flame-photometric (FPD)
- Pulse discharge detector in helium ionization or electron capture mode
- Micro-electronic capture detector (μECD) with a detection limit of 4.4 fg/s
Column thermostat and gas control
- The furnace range from 4 °C above ambient temperature to ≥450 °C, stability ±0.01 °C isothermically
- Cryogenic options: up to -80 °C with liquid nitrogen or up to -40 °C with CO2
- Electronic control of APC gases – up to 6 blocks and 18 channels for N2, He, H2, air and Ar/CH4 mixture
Sampling (injectors)
| Parameter | YL6500 GC |
|---|---|
| Injector, max. temperature | 450 °C |
| Nozzle injector, pressure | 0,001–100 psi |
| Nozzle injector, flow | 0.01–100 ml/min |
| Capillary Injector (split/splitless), Maximum Temperature | 450 °C |
| Flow division | up to 7,500:1 |
| Diameter of capillary columns | 0.05–0.53 mm |
Column thermostat
| Parameter | YL6500 GC |
|---|---|
| Temperature range | 4 °C above the ambient to ≥450 °C |
| Cryogenic cooling (option) | up to -80 °C (liquid nitrogen) / up to -40 °C (CO2) |
| Temperature stability | ±0.01 °C isothermically; ±0.1 °C in gradient |
Detectors.
| Parameter | YL6500 GC |
|---|---|
| PID (FID) | up to 450 °C, detection limit of 1.5 pg C/s, dynamic range 107 |
| Catarometer (TCD) | up to 400 °C, detection limit of 2.5 ng/ml |
| Thermoion Detector (NPD) | up to 400 °C, dynamic range >104 |
| Flame and Photometric Detector (FPD) | up to 300 °C (limit 250 °C), sensitive to sulfur and phosphorus |
| Pulse discharge detector | up to 400 °C, helium ionization or electron capture modes |
| Micro-detector of electronic capture (μECD) | up to 450 °C, detection limit 4.4 fg/s |
Gas control and SO
| Parameter | YL6500 GC |
|---|---|
| Electronic Gas Control (APC) | up to 6 blocks, 18 channels; N2, He, H2, air, Ar/CH4 |
| Software | YL-Clarity, conforming to 21 CFR Part 11, up to 4 chromatographs of 12 detectors in the network |