50 l/h
productivity
Micro Solutions / Life Science
Ultra-pure water type I: up to 50 l/h with resistance up to 18.2 MOM·cm
Ultra-pure water type I: up to 50 l/h with resistance up to 18.2 MOM·cm
UF, UV and VF configurations for endotoxin control and total organic carbon
aquapuri Essence+ is a type I water production system for sensitive analytical and Life Science tasks: HPLC, IX, AAS, ICP-MS with additional filter, cell culture, sequencing and PCR.
The system operates from pre-purified water type II or III and brings it to ultra-pure water type I. Standard, UF, UV and combined VF configurations are available: they allow you to choose the control of total organic carbon and / or removal of endotoxins depending on the method. water in the workplace.
Models 390, 391 UF, 392 UV and 393 VF have a performance of 50 l/h at 25 °C and a resistivity of ≥18.2 MOM·cm (conductivity ≤0.055 μSm/cm). TOC - ≤10 ppb for 390/391 and ≤5 ppb for 392/393; endotoxins <0.001 EU/ml for UF/VF; bacteria <0.01 CFU/ml, particles >0.22 μm missing, silicates <5 ppb, heavy metals < ppb1 for ICPb. The manufacturer requires a separate ICP filter.
productivity
resistance
TOS (392/393)
endotoxins (391/393)
front-dispenser
resistance monitoring and TOC
| Class water | Type I (standard) Type I (no pyrogens) Type I (without TOC) Type I (without pyrogens and TOC) |
|---|---|
| Performance at 25 °C | 50 l/h |
| Specific resistance at 25 °C | ≥18.2 MOM·cm (≤0.055 μS/cm) |
| TOS | ≤10 ppb ≤10 ppb ≤5 ppb ≤5 ppb |
| Endotoxins (pyrogens) | — <0.001 EE/ml — <0.001 EE/ml |
| Removal of bacteria | <0.01 CFU/ml |
| Particles larger than 0.22 microns | absent |
| Silicates | <5 ppb |
| Heavy metals | <1 ppb |
| Source water | ASTM type II or III water |
|---|
Removal of up to 99.99% of inorganic impurities to maintain 18.2 MOM·cm resistance
Removal of bacterial endotoxins (pyrogens), <0.001 EE/ml
Removal of trace organic substances and microorganisms, TOC <5 ppb
Finish removal of residual microorganisms and impurities larger than 0.22 μm
Additional filter for ICP-MS samples: removal of trace organic matter and ions