Yokogawa
High-content analysis system CellVoyager CV8000
High-performance cell screening: up to 4 cameras and 5 lasers, microplate per minute
CellVoyager CV8000 is Yokogawa’s automated HCA platform for high-performance imaging and quantitative analysis of living and fixed cell models.
Request a quotationWide-field confocal optics, up to four cameras, incubation and an optional robot dispenser support rapid multi-channel experiments, long-term time-lapse, 3D models and cell response studies to add reagents.
Up to 4 sCMOS cameras 2000×2000 pixels and up to 5 lasers 405/445/488/561/640 nm; up to 4 colors simultaneously in confocal mode. 6-1536 wells and subject slides microplates; incubation 35–40 °C. Measuring module 1280×895×1450 mm, weight 510 kg; Dry, water-immersion, phase-contrast and long-working-distance lenses are supported.
Throughput
- Up to 4 sCMOS cameras and 5 lasers capture a four-color 96-well microplate in about a minute.
- 6 to 1,536 holes and slide objects – from screening to single samples
- Laser and imaged autofocus support stable focusing in long runs
Working with living cells
- Incubation on stage supports 35–40 °C and a 5% CO2 supply with forced hydration
- Optional robot dispenser with disposable tips adds reagents during the experiment
- Up to 6 automatically switchable lenses, including water-immersion 20×/40×/60×
Analysis and data
- CellPathfinder program includes machine and deep learning, 3D analysis and label-free methods.
- Data is stored in 16-bit TIFF/PNG and CSV, LZW and ZIP compression is supported
- Optional barcode reader and paired workstations for control and analysis
Optics and detection
| Parameter | CV8000 |
|---|---|
| Confocal block | Nipkow dual-disc scanner with microlenses, pinhole 50 μm; 25 μm option and autodisc changer |
| Excitation wavelengths | 405/445/488/561/640 nm, up to 5 solid-state lasers (445 nm for FRET CFP/YFP); LED option 365 nm |
| Mode of shooting. | confocal, up to 4 colors at a time; light field and phase contrast 10×/20×; digital phase contrast |
| Camera. | sCMOS, 2000×2000 pixels, field of view 13×13 mm, up to 4 cameras |
| Objectives | dry 2×/4×/10×/20×; water immersion 20×/40×/60×; phase contrast 10×/20×; long working distance 20× |
Microplates and stage
| Parameter | CV8000 |
|---|---|
| Format of the sample | multi-well plates 6-1536 holes, subject slide |
| XY scene | travel X 120 mm, Y 80 mm, installation resolution 0.2 μm |
| Z-axis | range of 4 mm from the focal plane, resolution 0.1 μm |
| Autofocus. | laser-image |
| Incubation on stage | 35–40 °C; 5% CO2 supply with forced humidification; multigas unit option |
Automation and data
| Parameter | CV8000 |
|---|---|
| Robot dispenser | Optionally, disposable tips 96 or 384 |
| Barcode reader | optionally, 1D or 2D |
| Analysis | CellPathfinder: granularity, neurites, nuclear morphology and translocation, machine and deep learning, 3D analysis |
| Data formats | Images – 16-bit TIFF, PNG; numerical data – CSV and native format |
Dimensions and power supply
| Parameter | CV8000 |
|---|---|
| Dimensions of the measuring unit | 1280 × 895 × 1450 mm |
| Mass of the measuring unit | 510 kg |
| Working environment | 15–30 °C, 30–70% of relative humidity, without condensation |
| Power supply | measuring unit – 100–240 V, 50/60 Hz, up to 2 kVA; workstations – up to 1400 W and 1350 W |