Xian Tianlong
Automatic extraction workstation of NK PANA9600S
From scanning a sample to setting up four PCR systems in one run
Automated station for loading, extraction of NDT and installation of up to four PCR systems.
Request a quotationPANA9600S combines sample information reading, dosing, rotational magnetic extraction and PCR preparation. Four loading channels, reagent and consumable identification, LIS communication, refrigerated reagent chamber and separate zones reduce manual operations and transfer risk.
Capacity 1-96 samples; in one run, extraction of 96 samples and setting of four PCR systems is performed. Supported plasma, serum, whole blood, smears and urine; heating of lysis / elution 35-120 °C, camera PCR-reagents 4-15 °C, screen 12.1′′; 1370×810×890 mm, 220 kg.
Automation of the cycle
- One button starts scanning samples, loading, extraction of NA and setting PCR
- A full cycle for 96 samples for coronavirus is laid out in 50 minutes
- Four boot channels and the possibility of simultaneous setting of four PCR systems
Quality control
- Magnetic particle rotary extraction, smart droplet capture, strict zoning, air filtration and UV disinfection reduce the risk of cross-contamination
- Accurate dosing: at volume above 50 μl error ≤1.0%, repeatability CV ≤1.2%
- Barcode reagent scanning and visual recognition of consumables
Compatibility and integration
- Compatible with plasma, serum, whole blood, smears and urine
- Cooled chamber of PCR reagents with light protection (4-15 °C)
- Connection to LIS (Laboratory Information System)
Throughput
| Parameter | PANA9600S |
|---|---|
| Capacity | 1-96 samples |
| Processing for launch | extraction of 96 samples + setting of 4 PCR systems |
| Full cycle time (96 samples, COVID) | ~50 minutes |
| Types of specimens | plasma, serum, whole blood, swabs, urine |
| Booting channels | 4 |
Accuracy of dosing
| Parameter | PANA9600S |
|---|---|
| Accuracy/repeatability (<15 μL) | Δ≤2,0%; CV≤3,0% |
| Accuracy/repeatability (15–50 μL) | Δ≤1,5%; CV≤1,5% |
| Accuracy/repeatability (>50 μL) | Δ≤1,0%; CV≤1,2% |
| Monitoring of fluid levels | CapSense / Gas pressure sensor |
Regime and constructiveness
| Parameter | PANA9600S |
|---|---|
| Lysis and elution temperature | 35–120 °C |
| PCR reagent chamber | 4-15 °C, protection from light |
| Expenditures of PCR | strips 0.1/0.2 ml on 8 tubes, 96-well plates |
| Contamination control | zoning, HEPA-filtration, UV disinfection, capture of drops |
| Liaison to LIS | maintained |
| Display | 12.1 "touch" |
| Weight | 220 kg |
| Dimensions | 1370×810×890 mm (D×Sh×V) |