Micro Solutions

Analytical solutions for your laboratory

RU
Desktop nucleic acid extractor Libex

Xian Tianlong

Desktop nucleic acid extractor Libex

Automated nucleic acid extraction from 1–32 samples using magnetic particles

Benchtop automated extractor for simultaneous nucleic acid extraction from 1–32 samples.

Request a quotation

Libex automates lysis, particle binding, flushing and elution. Supports 96-well microplates and 6-strip tubes, local control and Android control over Wi-Fi, storage and editing protocols, UV disinfection and recovery after a power failure.

Performance 1-32, working volume 30-1000 μl, recommended sample 200 μl, the residue of magnetic particles ≤1%. The temperature of lysis and elution - from room to 120 °C; dimensions 435 × 440 × 445 mm, weight 31.5 kg, power 100-240 V.

Reliability of the result

  • Pre-filled magnetic particle sets require one step to start extraction
  • 32 SARS-CoV-2 RNA samples are processed in about 15 minutes (time depends on the reagent)
  • Magnetic particles not more than 1%

Management and convenience

  • Two control options: the keyboard of the device or Android application over Wi-Fi
  • Storage of up to 15 programs in the device and more than 500 programs in the application
  • Flexible creation, editing and deletion of protocols

Contamination control

  • Built-in UV disinfection module
  • Control system for cross-contamination of samples
  • Recovery after power failure at the user's choice

Throughput

ParameterLibex
Throughput1-32 sample
Workload30–1000 μl
Recommended sample size200 μl
Magnetic particle residue≤1%

Temperature and consumables

ParameterLibex
Lysis and elution temperatureroom-to-room 120°C
Compatible consumables96-well microplates, 6-strip tubes
Processing regimemultimode, variable-speed mixing

Design and power supply

ParameterLibex
Controlskeyboard of the device; Android application on Wi-Fi
ConnectionUSB
Network connectionWi-Fi
Weight31.5 kg
Dimensions435×440×445 mm (SH×G×V)
Power supplyAC 100–240 V, 50/60±1 Hz, 600 W
Operating temperature range10–30 °C
Humidity20–85 %