1–96
run-through
Micro Solutions / Life Science
Flexible extraction from 1 to 96 samples without loss of reagent when incompletely loaded

Flexible extraction from 1 to 96 samples without loss of reagent when incompletely loaded
6×16 module with 96-well microplates and 6-strip tubes
Flexible extraction of 1 to 96 samples with rotational stirring and contamination control.
The GeneRotex 96 works with 96-well microplates and 6-strip tubes, reducing reagent consumption when underloaded: a seven-inch screen, automatic door, library of up to 1,000 programs, negative pressure with HEPA and UV disinfection support routine high-flow processes.
Performance 1-96, reaction volume 30-1000 μl, recommended sample 200 μl, CV between positions ≤3%, magnetic particle residue ≤1%. Speed up to 3000 rpm; heating of lysis and elution from room temperature to 120 °C; 510 × 490 × 480 mm, 45 kg.
run-through
interpositional variability
magnetism
mass
module 6×16
saving of reagent in case of incomplete loading
| Throughput | 1-96 samples |
|---|---|
| Reaction volume | 30–1000 μl |
| Recommended sample size | 200 μl |
| Interpositional variability | CV≤3% |
| Magnetic particle residue | ≤1% |
| Compatible consumables | 96-well microplates, customized 6-strip tubes |
| Rotation speed | ≤ 3,000 rpm |
|---|---|
| Lysis and elution temperature | room-to-room 120°C |
| Mixing method | rotational (RMT) |
| Programming in memory | 1,000 |
| Display | 7' color sensory |
|---|---|
| Contamination control | Negative pressure, HEPA filter, UV disinfection |
| Connection | USB |
| Weight | 45 kg |
| Dimensions | 510×490×480 mm (Wh×D×V) |
| Power supply | AC 100–240 V, 50/60 Hz, 600 VA |
Pre-filled lysis, binding, flushing and elution reagents; compatible with 96-well microplates and 6-strip tubes