Xian Tianlong
Real-time PCR system Gentier 96E/96R
Quality control of NDT and libraries before sequencing: 96 holes scanned in 7 seconds
96-position real-time PCR platform Gentier for quality control NC and libraries.
Request a quotationThe abbreviated designation "96ER" from the source file is matched to the official family of Gentier 96E/96R. Before publication of the catalog number, it is necessary to confirm the modification; the family supports 4- or 6-channel detection, gradient and quantitative analysis.
Official Gentier 96E/96R: 96 positions, scanning in 7 s, heating ≥6.1 °C/s, cooling ≥5.0 °C/s, accuracy ≤0.1 °C and uniformity ±0.1 °C. Supported by 0.2 ml microplates, strips and single tubes.
Speed
- Full fluorescence scan of all 96 holes takes 7s
- Heating rate ≥6.1 °C/s, cooling rate ≥5.0 °C/s on Peltier cells
- Automatic recovery of the experiment after a power failure
precision
- Unit temperature uniformity ±0.1 °C, maintenance accuracy ≤0.1 °C
- Linearity |r| >0.999, repeatability Ct CV ≤0.5%
- Gradient block with a range of 1–40 °C for optimization of protocols
Flexibility of application
- Two control options: 10.4-inch touch screen or PC software
- Support for gradient, long and Touch Down PCR
- Up to 1000 results are stored in the device memory, data exchange via USB
Throughput
| Parameter | Gentier 96E | Gentier 96R |
|---|---|---|
| Number of positions | 96 | 96 |
| Number of fluorescent channels | 6 | 4 |
| All-hole scanning. | 7 | 7 |
Temperature control
| Parameter | Gentier 96E | Gentier 96R |
|---|---|---|
| Method of heating | peltier | |
| Max, heating speed | ≥ 6.1 °C/s | |
| Max, cooling speed | ≥ 5.0 °C/s | |
| Uniformity | ±0,1 °C | |
| precision | ≤0,1 °C | |
| Gradient range | 1-40 °C, 12 rows |
Optical system and software
| Parameter | Gentier 96E | Gentier 96R |
|---|---|---|
| Light source | LED, top-up excitation | |
| Detector | photodiode, top scan. | |
| Linearity | |r| >0,999 | |
| Repeatability (Ct) | CV ≤0,5 % | |
| Functions of analysis | absolute and relative quantitative estimation, SNP analysis, analysis of melting curves |
Technical parameters
| Parameter | Gentier 96E | Gentier 96R |
|---|---|---|
| Controls | 10.4-inch touchscreen or PC (Win7/Win10) | |
| Data storage | Up to 1000 results, exchange via USB | |
| Consumables | 0.2 ml 96-well microplates, 8-strips, single tubes | |
| Weight | 30 kg | |
| Dimensions | 355×475×484 mm | |
| Power supply | AC 100–240 V, 50–60 Hz, 900 VA |