Micro Solutions Life ScienceRU

Micro Solutions / Life Science

The gel documentation system omniDOC

Modular documentation of DNA, RNA and protein gels: 5 MP camera with interchangeable filters

Product overview

Modular documentation of DNA, RNA and protein gels: 5 MP camera with interchangeable filters

with a retractable UV transilluminator of 312 nm and an optional blue LED module

Cleaver Scientific omniDOC is a modular gel documentation system for DNA, RNA and protein samples. The 5MP camera, UV transilluminator and interchangeable optical modules give the lab a single working tool for fluorescent and colorimetric recording.

The base system combines a closed camera, a 312 nm retractable UV transilluminator and white EPI illumination for positioning. Interchangeable filters adapt the device to EtBr, SYBR Green, runSAFE and other fluorophores. Optional blue LED module allows you to work with safe dyes, and a white light table with colorimetric gels and visible lights. The software does the capture and basic analysis of images; the data is stored on a connected PC or laptop.

Monochrome CMOS camera 1/2.5 inches, resolution 5 MP (up to 2592×1944), 12-bit image with 4096 gray levels. lens with a focal length of 5 mm and an aperture of F1.2; sample area up to 21×26 see UV transilluminator 312 nm size 21×26 cm with six lamps of 8 W. Standard filter EtBr 620 nm; optionally 520, 560 and 580 nm and neutral filter: A blue 470 nm EPI excitation module and a white 21x26 cm light table are available as options. USB connection to a PC, camera dimensions 410×405×570 mm, weight about 25 kg, power 110–230 V AC; a door lock switch turns off the UV when opened.

Key figures

5 MP

camera

312 nm

UV transilluminator

21x26 cm

field

25kg

camera-room

Applications

Models and configurations

Product details

Place in the system line

Comparison with microDOC, gelLITE, gelONE

Modules and specifications

camera, transilluminator, options

Shooting.

Light sources

Security and software

Camera.

TypeCMOS 1/2.5” monochrome
Permission5 MP (up to 2592×1944)
Depth of image12 bits, 4096 gray gradations
ObjectiveF1.2, focal length 5 mm

Transillumination

UV transilluminator312 nm, 21x26 cm, 6 lamps of 8 W
Camera filter620 nm EtBr (standard); optionally 520/560/580 nm
Blue EPI module (option)470 nm
White light table (option)21x26 cm

Dimensions and power supply

Field of photography21x26 cm
Dimensions (dark room)410×405×570 mm
Weightabout 25kg
Power supply110-230 In AC

System modifications

omniDOC SAFE

base system + blue LED module and filters 520/560/580 nm

omniDOC PRO

Basic system + white light table and ND filter

omniDOC PRO SAFE

base system + blue LED module, filters 520/560/580/ND and light table

Options and filters

White light table

For colorimetric gels and visible blots, 21x26 cm

Blue EPI LED module

Excitation of 470 nm for safe dyes

Related equipment types

Classification

Documents

Manufacturer sources