Frontal autoclaves 100-400 liters for sterilization of media, dishes and waste decontamination

Priorclave lab autoclaves with front loading

Product overview

Frontal autoclaves 100-400 liters for sterilization of media, dishes and waste decontamination

cylindrical camera, Tactrol 3 control and thermally locked Quickseal door

Priorclave front autoclaves are a versatile lineup for sterilizing nutrient media, laboratory utensils, tools and waste in research, training and manufacturing laboratories. The cylindrical chamber provides efficient steam circulation and helps reduce water and energy consumption compared to large rectangular sterilizers.

The equipment performs programmable cycles of steam sterilization with preparation of the medium, automatic removal of air, exposure, cooling and temperature maintenance depending on the chosen configuration. Front loading is convenient for shelves, containers and large lots of the same type of laboratory dishes. The task, and the Smart configurations add vacuum, drying and advanced work with complex and porous boots. Tactrol system manages the time and temperature of the cycle; door locks Quickseal do not allow you to open the camera at dangerous temperature or pressure.

The current line includes models EH/SH 100, 150, 200, 320 and 400 with a useful volume of 100-400 liters. the Working range is up to 138 °C and 2.4 bar. 100/150/200 liters have a diameter of 500 mm and a depth of 535/740/1025 mm; models 320/400 l - a diameter of 630 mm and a depth of 1025/1200 mm. Heating: internal electric or from an external source of steam; power up to 7-10.5 The 100-200 liter and 18 kW models have 320-400 liter cameras and doors made of 316L stainless steel. The standard is Tactrol 3 programmable control, automatic free steaming, low-water sensor, Media Warming and Delayed Start modes, rollover shelves and antimicrobial treatment of exterior panels. Specific vacuum, cooling and recording options are selected for the boot type.

Key figures

100-400 l

volume-range

138 °C

temperature

2.4 bar

max.

CE

PED 2014/68/EU

Applications

Models and configurations

Product details

Autoclaves 100–200 l

front loading, cylindrical camera

Autoclaves 320-400 l

front-loading

Safety

Cycle management

Operation

Camera and loading

Useful volume100 l.
150.
200.
320 l
400 l.
Diameter of the camera500 mm
500 mm
500 mm
630 mm
630 mm
The depth of the chamber535 mm
740 mm
1025 mm
1025 mm
1200 mm
Load height860 mm
860 mm
860 mm
725 mm
725 mm
Type of loadingfrontal

Sterilization regimen

Operating rangeup to 138 °C, 2.4 bar

Heating and power supply

heat sourceinternal electric or external steam
Electrical power7 kW (30 A, 1 phase) / 10.5 kW (15 A, 1 phase)
7 kW (30 A, 1 phase) / 10.5 kW (15 A, 1 phase)
7 kW (30 A, 1 phase) / 10.5 kW (15 A, 1 phase)
18 kW, 30 A, 3 phases
18 kW, 30 A, 3 phases

Dimensions and mass

Dimensions (W×D×H)735 x 988 x 1529 mm
735 x 1215 x 1529 mm
735 x 1469 x 1529 mm
760 x 1310 x 1525 mm
760 x 1575 x 1525 mm
Mass without loading190 kg
208 kg
236 kg
300 kg
350 kg

Materials and conformity

Camera material and doors316L stainless steel
Pressure vessel warranty10 years.
ConformityCE/UKCA, PED 2014/68/EU, PD5500:2000 Cat.3, BS2646-2021, ISO 9001; EMC BS EN 61000 and BS EN 61326

Additional accessories

Booting cassettes with adjustable separators

Loading during sterilization of small items

Stabilized loading baskets

Additional carts to increase the useful load

Replaceable load temperature sensors with cables

Replacement of “floating” temperature probes to control the cycle

Monitoring and control options

Cycle data printer

Printing the sterilization protocol

Monitoring of loading temperature

Control of temperature directly in loading

Monitoring of pressure in the chamber

Additional pressure control during the cycle