100-400 l
volume-range
Micro Solutions / Life Science
Frontal autoclaves 100-400 liters for sterilization of media, dishes and waste decontamination

Frontal autoclaves 100-400 liters for sterilization of media, dishes and waste decontamination
cylindrical camera, Tactrol 3 control and thermally locked Quickseal door
Priorclave front autoclaves are a versatile lineup for sterilizing nutrient media, laboratory utensils, tools and waste in research, training and manufacturing laboratories. The cylindrical chamber provides efficient steam circulation and helps reduce water and energy consumption compared to large rectangular sterilizers.
The equipment performs programmable cycles of steam sterilization with preparation of the medium, automatic removal of air, exposure, cooling and temperature maintenance depending on the chosen configuration. Front loading is convenient for shelves, containers and large lots of the same type of laboratory dishes. The task, and the Smart configurations add vacuum, drying and advanced work with complex and porous boots. Tactrol system manages the time and temperature of the cycle; door locks Quickseal do not allow you to open the camera at dangerous temperature or pressure.
The current line includes models EH/SH 100, 150, 200, 320 and 400 with a useful volume of 100-400 liters. the Working range is up to 138 °C and 2.4 bar. 100/150/200 liters have a diameter of 500 mm and a depth of 535/740/1025 mm; models 320/400 l - a diameter of 630 mm and a depth of 1025/1200 mm. Heating: internal electric or from an external source of steam; power up to 7-10.5 The 100-200 liter and 18 kW models have 320-400 liter cameras and doors made of 316L stainless steel. The standard is Tactrol 3 programmable control, automatic free steaming, low-water sensor, Media Warming and Delayed Start modes, rollover shelves and antimicrobial treatment of exterior panels. Specific vacuum, cooling and recording options are selected for the boot type.
volume-range
temperature
max.
PED 2014/68/EU
front loading, cylindrical camera
front-loading
| Useful volume | 100 l. 150. 200. 320 l 400 l. |
|---|---|
| Diameter of the camera | 500 mm 500 mm 500 mm 630 mm 630 mm |
| The depth of the chamber | 535 mm 740 mm 1025 mm 1025 mm 1200 mm |
| Load height | 860 mm 860 mm 860 mm 725 mm 725 mm |
| Type of loading | frontal |
| Operating range | up to 138 °C, 2.4 bar |
|---|
| heat source | internal electric or external steam |
|---|---|
| Electrical power | 7 kW (30 A, 1 phase) / 10.5 kW (15 A, 1 phase) 7 kW (30 A, 1 phase) / 10.5 kW (15 A, 1 phase) 7 kW (30 A, 1 phase) / 10.5 kW (15 A, 1 phase) 18 kW, 30 A, 3 phases 18 kW, 30 A, 3 phases |
| Dimensions (W×D×H) | 735 x 988 x 1529 mm 735 x 1215 x 1529 mm 735 x 1469 x 1529 mm 760 x 1310 x 1525 mm 760 x 1575 x 1525 mm |
|---|---|
| Mass without loading | 190 kg 208 kg 236 kg 300 kg 350 kg |
| Camera material and doors | 316L stainless steel |
|---|---|
| Pressure vessel warranty | 10 years. |
| Conformity | CE/UKCA, PED 2014/68/EU, PD5500:2000 Cat.3, BS2646-2021, ISO 9001; EMC BS EN 61000 and BS EN 61326 |
Loading during sterilization of small items
Additional carts to increase the useful load
Replacement of “floating” temperature probes to control the cycle
Printing the sterilization protocol
Control of temperature directly in loading
Additional pressure control during the cycle