UV/Visible Spectrophotometer MicroDigital Ubi-490

Micro Digital

UV/Visible Spectrophotometer MicroDigital Ubi-490

Compact single-beam divet UV/visible spectrophotometer for routine photometry and water analysis

Ubi-490 is a compact single-beam UV/visible MicroDigital spectrophotometer for routine photometry, quantitative analysis and spectral studies of liquid and light-transmitting samples.

Request a quotation

It measures the absorption and transmission of samples in ditches, and with additional holders it works with microvolumes of less than 500 μl, tubes, optical films and object slides. It supports photometric and quantitative analysis, spectrum scanning, multiwavelength and kinetic measurements. In positioning, the eight-millimeter aluminum base supports the stability of the optical system, and the instrument interface simplifies the settings.

Photometric system - single beam. wavelength range 190-1100 nm, spectral band 2 nm, accuracy of wavelength setting ± 0.5 nm, reproducibility 0.3 nm Photometric accuracy ±0.3%T, repeatability 0.2%T; display range 0-200%T and −0.3-3.0 A. Stability 0.002 A/h at 500 nm, scattered light ≤0.05%T at 220 and 360 nm. Sources - Deuterium and tungsten-halogen lamps, detector - silicon photodiode. output data via USB; power 220 V, 50 Hz. Dimensions 450×380×180 mm, weight 18 kg.

Measurements

  • Measurement of absorption and transmission of samples in ditch
  • Additional holders for microvolumes less than 500 μl, tubes, films and subject slides
  • Support for photometric and quantitative analysis, spectrum scanning, multiwavelength and kinetic measurements

Accuracy and stability

  • The screw drive increases positioning accuracy
  • Eight-millimeter aluminum base supports optical system stability

Application

  • The manufacturer separately indicates the use in water analysis
  • Determination of pesticide residues

Optical system

ParameterUbi-490
Scheme.single-beam (single beam)
Wavelength range190-1100 nm
Spectral stripe2 nm
Accuracy of wavelength setting± 0.5 nm
Reproduction of wavelength0.3 nm
Scattered light≤0.05%T at 220 and 360 nm

Photometry

ParameterUbi-490
Photometric accuracy±0,3%T
Repeatability0,2%T
Display range0–200%T; −0,3–3,0 A
Stability0.002 A/h at 500 nm

Source and detector

ParameterUbi-490
Radiation sourcesdeuterium and tungsten-halogen lamp
Detectorsilicon-photodiode

power supply and dimensions

ParameterUbi-490
Data interfaceUSB
Power supply220 V, 50 Hz
Dimensions450×380×180 mm
Weight18 kg