5
slide-tray
Micro Solutions / Life Science
Compact fluorescent 5-slide WSI scanner for immunofluorescence and FISH

Compact fluorescent 5-slide WSI scanner for immunofluorescence and FISH
up to 10 channels, cooled sCMOS camera, with light-field mode
A compact 5-slide fluorescent WSI scanner for immunofluorescence, FISH and research morphology, up to 10 channels and a refrigerated sCMOS camera give the lab full multi-channel desktop imaging.
KF-FL-005 performs area fluorescent and linear light field scans, supports configurations on 3, 4, 6, 8 or 10 fluorescent channels, Z-stack up to 10 layers and fusion of focal planes. Fluorescence results are stored in KFBF and QPTIFF, light field in KFB, SVS, DCM and TIFF.
Capacity: 5 slides. Fluorescence: cooled sCMOS camera, 0.25/0.125 μm per pixel, up to 6 minutes for a 10×10 mm area with 3 filters and up to 12 minutes with 6 filters. Light field: 20×/40×, 15 seconds on slide, up to 72 slides/h at 20×. Field of view 1 mm. Dimensions 470×510×570 mm, weight 53 kg, power 150 VA.
slide-tray
fluorescent channels
slide/h (light field)
power-use
fluorescent and light-field scanning
470 x 510 x 570 mm, 53 kg
| Capacity | 5 slide |
|---|---|
| Fluorescent scanning speed | up to 6 minutes (3 filters) / 12 minutes (6 filters) per 10 × 10 mm area |
| Light-field scanning speed | from 15 s per slide, up to 72 slides / h at 20 × |
| Camera. | sCMOS-cooled |
|---|---|
| Channels | up to 10 (configurations 3/4/6/8/10) |
| Permission | 0.25 μm/pixel (20×) / 0.125 μm/pixel (40×) |
| Field of vision | 1 mm |
| Method of scanning | Linear (light field) + area (fluorescence) |
|---|---|
| Objective | OLYMPUS, apochromatic, 20×/40× |
| Z-stack | 10 layers |
| Formats (fluorescence) | KFBF, QPTIFF |
|---|---|
| Formats (light field) | KFB, SVS, DCM, TIFF |
| Slide marking | barcode/QR code |
| Dimensions (L×W×H) | 470 x 510 x 570 mm |
|---|---|
| Weight | 53 kg |
| Power consumption | 150 VA |