Micro Solutions Life ScienceRU

Micro Solutions / Life Science

Biological Safety Box of the HFsafe LC B2 Series

Type II B2 Class with 100% air removal from the work area

Biological Safety Box of the HFsafe LC B2 Series

Product overview

Type II B2 Class with 100% air removal from the work area

for operation without recirculation of air inside the laboratory, sizes 900-1800LCB2

The Heal Force HFsafe LC B2 is a Type II B2 biosafety box with one hundred percent outdoor air removal from the work area, designed for tasks that require a hard connection to the exhaust system and eliminate air recirculation inside the laboratory.

Boxing protects the operator, sample and environment by maintaining controlled incoming and downstream flows and diverting all treated air outward without recirculation.Flow-saFe balances air speeds, independent sensors monitor flows, and LCD shows air condition, glass position, temperature, filter resource and time Programmable UV disinfection, protective locks, constant exhaust flow and backflow protection are provided; the working area is made of 304 stainless steel.

Class II, type B2, air circuit - 100% exhaust without recirculation. The official Heal Force technical page indicates a downstream velocity of 0.30 m/s and an incoming flow of 0.53 m/s. Used ULPA filter of borosilicate fiberglass; the change in flow rate remains less than 10% when the filter resistance changes by 50%. NSF 49, EN 12469N and CFDA certificates; AC 220 B, 50 Hz power; Exact size (900LCB2/1200LCB2/1500LCB2/1800LCB2), exhaust system performance and external dimensions are selected for a specific supply and a separate table is not published on the current official page.

Key figures

II B2

protection

100%

recirculation

NSF49 / EN12469N

certificate

Applications

Models and configurations

Product details

Biosafe Boxing HFsafe LC B2

100% air extraction

Typical series

900–1800LCB2

Air circuitry

Indication and safety

Design

Protection class

Type of boxingClass II, type B2
Air circuitry100% exhaust (100% exhaust), without recirculation

Airflow

Downstream velocity0.30 m/s
Incoming flow rate0.53 m/s

Filtration

FilterULPA from borosilicate fiberglass
Stability of flowChange less than 10% when filter resistance changes by 50%

Power supply and certificates

Power supplyAC 220 V, 50 Hz
CertificatesNSF 49, EN 12469N, CFDA