Heal Force
The HFsafe-1200LC Biological Safety Box
Compact size of HFsafe LC Class II A2 series for limited-area laboratories
HFsafe-1200LC is a compact version of Heal Force Class II Type A2 biosafety boxes for laboratories with a limited work area.
Request a quotationIt uses the common HFsafe LC architecture: Flow-saFe system for automatic flow balance, negative pressure in double walls, independent control of incoming and downstream streams, ULPA filtering, LCD monitoring and programmable UV decontamination.
Class II, type A2. External dimensions 1340×790×2130 mm, internal dimensions of the working area 1250×575×625 mm Incoming flow speed 0.53±0.025 m/s, downwards - 0.35±0.025 m/s, noise level not more than 63 dB, illumination of the working area not less than 1000 lux Maximum power consumption 1600 W with power supply 220 V, 50 Hz. NSF 49 and EN 12469N.
Control of airflows
- Flow-saFe automatically supports the ratio of incoming and downstream streams
- Negative pressure in double walls, independent control of incoming and downstream flows
Operation
- ULPA-filtration of inflow and exhaust air
- LCD parameter monitoring and programmable UV decontamination
Protection class
| Parameter | 1200LC |
|---|---|
| Type of boxing | Class II, type A2 |
Dimensions
| Parameter | 1200LC |
|---|---|
| External dimensions (W×D×H) | 1340×790×2130 mm |
| Internal dimensions of the working area (WxGxB) | 1250×575×625 mm |
Airflow
| Parameter | 1200LC |
|---|---|
| Incoming flow rate | 0.53 ± 0.025 m/s |
| Downstream velocity | 0.35 ± 0.025 m/s |
| Noise level | ≤63 dB |
| Illumination of the work area | ≥ 1000 lux |
Filtration
| Parameter | 1200LC |
|---|---|
| Filter | ULPA |
| Filter efficiency | 99.999% for particles 0.1–0.2 μm |
Power supply and certificates
| Parameter | 1200LC |
|---|---|
| Power supply | 220 V, 50 Hz |
| Maximum power consumption | 1600 watts |
| Certificates | NSF 49, EN 12469N, CFDA |