Gram Bioline
Ultra-temperature freezer BioUltra UL570
Long-term storage of biosamples at 86 °C: cascading cooling and vacuum insulation
Gram BioLine BioUltra UL570 is an ultra-low-temperature freezer developed and manufactured in Denmark for the long-term storage of valuable biological samples at temperatures up to -86 °C.
Request a quotationA cascade refrigeration system, vacuum thermal insulation panels and two independently open interior doors help to keep samples stable. Five perforated stainless steel shelves are standard; two 24.5 mm diameter ports allow external sensors to be installed. Supports programmable acoustic alarms, event logging, high/low temperature indication, E-sensor, reading correction and dry contact. Hybrid air-water capacitor is available.
BioUltra UL570 operates in the range of −86...−60 °C at an ambient temperature of +10 ... + 30 °C. Total volume - 570 liters; dimensions - 856 × 979 × 1986/1996 mm, weight without packaging and options - 257 kg Capacity reaches 384 boxes 50 mm high or 192 boxes 75 or 100 mm high. 70 mm insulation with vacuum panels, cascading refrigerants are used R290/R170/R601, power supply 230 V, 50 Hz, 16 A; power consumption 10.5 kWh/day, noise level 46.5 dB(A). The cabinet meets the internal and external requirements of ATEX Category 3 Zone 2.
Stability at -86 °C
- Gram BioLine Cascade Refrigeration System is specially designed for ultra-low temperatures and provides high efficiency and reliability.
- 70 mm insulation with vacuum thermal insulation panels (VIP) of the latest generation
- Two independently opened insulated interior doors maintain stable conditions even with the external door ajar, removable for cleaning.
Capacity and access
- 5 perforated stainless steel shelves are standard, adjustable in 12.5 mm increments
- Capacity: up to 384 boxes of 50 mm, 192 boxes of 75 mm, or 192 boxes of 100 mm
- Two 24.5mm ports for external sensors and an ergonomic handle for camera access
- Multifunctional leveling rollers for positioning with leg stability
Controls and options
- Gram controller with non-potential contact, E-sensor, programmable acoustic temperature alarm, alarm recording, high/low temperature indication and reading correction function
- Optional hybrid water-air capacitor reduces heat and noise and provides redundancy in case of interruptions in the supply of technical water
- IQ/OQ documentation is available for download from the manufacturer’s website
Temperature and regimen
| Parameter | UL570 |
|---|---|
| Temperature range | −86…−60 °C |
| Permissible room temperature | +10…+30 °C |
| Controller | Gram Control Unit: non-potential contact, E-sensor, programmable acoustic temperature alarm, alarm recording, high/low temperature indication, reading correction function |
Dimensions and capacity
| Parameter | UL570 |
|---|---|
| External dimensions (ExxGxV) | 856 x 979 x 1986/1996 mm |
| Total volume | 570 l |
| Mass without options | 257 kg |
| Capacity, 50 mm boxes | 384 pcs (stands 4×4) / 38,400 tubes |
| Capacity, boxes 75 mm | 192 pcs (stands 3×4) / 19,200 tubes |
| Capacity, 100 mm boxes | 192 pcs (stands 2×4) / 19,200 tubes |
Refrigeration system and energy consumption
| Parameter | UL570 |
|---|---|
| Type of system | static, cascading refrigeration system |
| Refrigerant | Step 1 – R290, Step 2 – R170/R601 |
| Energy consumption | 10.5 kWh/day (empty cabinet, set −80 °C, +20 °C of the environment) |
| Noise level | 46.5 dB(A) |
| Isolation. | 70 mm, without HFC propellant cyclopentane, vacuum panels |
Design
| Parameter | UL570 |
|---|---|
| Materials of the inner chamber | white painted steel, interior of doors and bottom panel - stainless steel |
| Case material | white-coloured steel or stainless steel |
| Ground | leveling |
| Power supply | 230 V, 50 Hz, 16 A |
| Ports of access | 2 x 24.5 mm |
| Sensors. | PT-10 000 |