Micro Solutions Life ScienceRU

Micro Solutions / Life Science

Ultra-temperature freezer BioUltra UL570

Long-term storage of biosamples at 86 °C: cascading cooling and vacuum insulation

Ultra-temperature freezer BioUltra UL570

Product overview

Long-term storage of biosamples at 86 °C: cascading cooling and vacuum insulation

Two independent interior doors and up to 384 boxes

Gram BioLine BioUltra UL570 is an ultra-low-temperature freezer developed and manufactured in Denmark for the long-term storage of valuable biological samples at temperatures up to -86 °C.

A cascade refrigeration system, vacuum thermal insulation panels and two independently open interior doors help to keep samples stable. Five perforated stainless steel shelves are standard; two 24.5 mm diameter ports allow external sensors to be installed. Supports programmable acoustic alarms, event logging, high/low temperature indication, E-sensor, reading correction and dry contact. Hybrid air-water capacitor is available.

BioUltra UL570 operates in the range of −86...−60 °C at an ambient temperature of +10 ... + 30 °C. Total volume - 570 liters; dimensions - 856 × 979 × 1986/1996 mm, weight without packaging and options - 257 kg Capacity reaches 384 boxes 50 mm high or 192 boxes 75 or 100 mm high. 70 mm insulation with vacuum panels, cascading refrigerants are used R290/R170/R601, power supply 230 V, 50 Hz, 16 A; power consumption 10.5 kWh/day, noise level 46.5 dB(A). The cabinet meets the internal and external requirements of ATEX Category 3 Zone 2.

Key figures

−86 °C

temperature

384

50 mm boxes (4×4 racks)

10.5 kWh/day

power consumption

ATEX

Cat.3 Zone 2 inside and out

Applications

Models and configurations

Product details

Ultra-temperature freezer BioUltra UL570

5 perforated stainless steel shelves complete

Internal doors and access ports

2 independent interior doors, 2 ports 24.5 mm

Stability at -86 °C

Capacity and access

Controls and options

Temperature and regimen

Temperature range−86…−60 °C
Permissible room temperature+10…+30 °C
ControllerGram Control Unit: non-potential contact, E-sensor, programmable acoustic temperature alarm, alarm recording, high/low temperature indication, reading correction function

Dimensions and capacity

External dimensions (ExxGxV)856 x 979 x 1986/1996 mm
Total volume570 l
Mass without options257 kg
Capacity, 50 mm boxes384 pcs (stands 4×4) / 38,400 tubes
Capacity, boxes 75 mm192 pcs (stands 3×4) / 19,200 tubes
Capacity, 100 mm boxes192 pcs (stands 2×4) / 19,200 tubes

Refrigeration system and energy consumption

Type of systemstatic, cascading refrigeration system
RefrigerantStep 1 – R290, Step 2 – R170/R601
Energy consumption10.5 kWh/day (empty cabinet, set −80 °C, +20 °C of the environment)
Noise level46.5 dB(A)
Isolation.70 mm, without HFC propellant cyclopentane, vacuum panels

Design

Materials of the inner chamberwhite painted steel, interior of doors and bottom panel - stainless steel
Case materialwhite-coloured steel or stainless steel
Groundleveling
Power supply230 V, 50 Hz, 16 A
Ports of access2 x 24.5 mm
Sensors.PT-10 000

Options

Hybrid water-air condenser

Reduces heat and noise levels, provides redundancy in case of interruptions in the supply of technical water (requires the connection of technical water)

Related equipment types

Related products

Classification

Product images

Documents

Manufacturer sources