180
tests/h
Micro Solutions / Life Science
Compact automated clinical chemistry for small laboratories: up to 180 tests per hour, or 360 with ISE

Compact automated clinical chemistry for small laboratories: up to 180 tests per hour, or 360 with ISE
optional ISE module (Na, K, Cl, Li) and primary sampling
A compact automated biochemistry analyzer for small and medium-sized laboratories. XL 180 combines routine and specialized chemistry, immunoturbidimetry, and an optional ISE module in one benchtop system.
Random-access operation supports STAT samples, automatic dilution and reruns. Sample and reagent barcode reading, refrigerated reagents, liquid-level sensing, and a separate mixer reduce manual work.
Up to 180 photometric tests per hour, or 360 with optional ISE. Minimum reagent volume: 180 µL. The static photometer uses 8 wavelengths from 340 to 700 nm. Primary tubes, disposable or washable cuvettes, and several calibration methods are supported.
tests/h
test-hours
wavelengths
mass
180 tests/h, up to 360 with ISE
Intuitive interface, control of analysis
| Throughput | Up to 180 tests/h (photometry); up to 360 tests/h with optional ISE |
|---|---|
| Reaction volume | 180 μl |
| Reagent dosing | 10–300 μl, step 1 μl |
| Measurement regimes | 1-Point, 2-Point, Rate-A, Rate-B, direct potentiometry (option) |
| Photometer | multiwave |
|---|---|
| Wavelengths | 8:340, 405, 505, 546, 578, 600, 660, 700 nm |
| cuvettes | PMMA rotor, 120 cuvette, optical path length 5 mm |
| Reagent positions | 25 fixed + 15 flexible (reagent/sample) |
|---|---|
| Number of samples | Up to 25 at the same time (any position is under STAT) |
| Type of sample | primary tube 7 ml (14.5×84 mm) or 5 ml (13×75 mm), cuvette 2 ml |
| Programmable parameters | up to 42 photometric tests + 4 ISEs (Na, K, Cl, Li, option) |
| Power supply | 220 V ±10 % or 110 V ±10 %, 50/60 Hz, 600 VA (with PC and printer) |
|---|---|
| Water consumption | ~1.8 l/h |
| Dimensions of the working area | 675×450×300 mm |
| Massa | 65 kg |
Map of test resource for 500,000 tests for a period of 1 year