Micro Solutions Life ScienceRU

Micro Solutions / Life Science

Identification system and MIKROLA ABCH

Reading, identification and interpretation of microbial sensitivity according to EUCAST and CLSI

Identification system and MIKROLA ABCH

Product overview

Reading, identification and interpretation of microbial sensitivity according to EUCAST and CLSI

Photometer, software and equipment for MIKROLATEST and SENSILATEST panels

MIKROLA is a modular system for automated reading, identification of microorganisms and interpretation of antibiotic sensitivity.

The complex combines the LISASCAN EM or MULTISKAN EX photometer, MIKROB-2 / MIKROB AUTOMAT, MIKROLATEST ID and SENSILATEST tests, as well as the densitometer and dosing facilities. The panel results can be read automatically, interpreted by EUCAST / CLSI and transmitted to the laboratory system.

The composition depends on the configuration: microplate reader, specialized software, ID/AST panels and DENSILAMETER; recommended additional modules are VORTEX, MIKROLASTEPPER/MULTISTEPPER and VIAFLO Vision. The specific reader model and completeness must be fixed in a commercial configuration.

Key figures

8

measuring-channel

EUCAST/CLSI

expert interpretation

0,001 Abs

photometric resolution

Applications

Models and configurations

Product details

The MIKROLA complex

reader, software and auxiliary equipment

Reading and software

Auxiliary equipment

Photometer LISASCAN EM

Optical system8-channel fiber optics (8 vertical channels)
Light sourcehalogen lamp 13.8 V/50 W
DetectorSi-photodiode
Light filters405, 450, 492, 578, 630, 700 nm
Photometric resolution0,001 Abs
Measurement range0–3 Abs
precisionless than 1% at 2.5 OD
Conditions of operation20–40 °C
Dimensions502 x 369 x 216 mm
Weight7 kg

Software

MIKROB-AUTOMATReading and archiving of MIKROLATEST ID, BP and MIC results; transferring data to MIKROB2
MIKROB2EUCAST/CLSI expert system, patient registers, statistics and epidemiological analysis

Auxiliary devices

Densi-La-Meter IIoptical determination of turbidity according to MacFarland, range 0.0-15.0, calibration at 3 points 0.5-9.0
VIAFLO Visionsingle-channel pipette 50–1250 μl

Readers and POs

LISASCAN EM

Universal 8-channel photometer for microbiology and immunology

MIKROB-AUTOMAT

Software for reading and managing the ID/BP/MIC panels

MIKROB2

EUCAST/CLSI expert interpretation system with statistics

Auxiliary equipment

DENSILAMETER II

Densitometer for determining bacterial inoculate by McFarland

vortex

homogenization

VIAFLO Vision, pipette 50–1250 μl

Electronic dosing for MIKROLATEST and SENSILATEST panels

Related equipment types

Related products

Classification

Product images

Documents

Manufacturer sources