Micro Solutions

Analytical solutions for your laboratory

RU
H 360 hematology analyzer

Erba

H 360 hematology analyzer

3-part differential CBC from just 9 µL of blood, at 60 tests per hour

H 360 is an automated 3-part differential hematology analyzer delivering 60 tests per hour with a sample volume of only 9 µL.

Request a quotation

It produces a complete blood count with a 3-part white-cell differential, extended platelet indices, and Mentzer/RDWI screening indices. A touchscreen, built-in printer, RFID reagent tracking, anti-clogging features, and LIS connectivity support a compact workflow.

24 parameters and 3 histograms. Impedance measurement for RBC, PLT, WBC, and differential counts; cyanide-free colorimetry for hemoglobin. Throughput: 60 tests per hour. Sample volume: 9 µL whole blood or 20 µL in prediluted/capillary mode. Stores 50,000 results. Dimensions: 364 × 477 × 417 mm; weight: 25 kg. 4 USB ports and LAN.

Accuracy and minimum sample size

  • 9 μl of whole blood per test, 20 μl in prediluted or capillary mode
  • Electrical impedance for RBC/PLT/WBC/DIFF, cyanide-free colorimetry for Hb
  • Mentzer and RDWI screening indexes for differential diagnosis of anemia
  • Guaranteed Dilution Accuracy by Automatic Diluent Delivery (ADDM)

Ease of operation

  • 10.4 touch screen, analysis in one click
  • Full traceability of RFID reagents
  • Built-in thermal printer, reports with histograms
  • Technology of protection against aperture clogging

Data storage and communication

  • Memory on 50,000 results with graphs
  • 4 USB ports and LAN port, connection to LIS
  • Possibility to include microscopic parameters (sticks, blasts, etc.) in the report

Measurement

ParameterH 360
Parameters24: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PCT, PDW-CV, PDW-SD, P-LCR, P-LCC, Mentzer and RDWI, Lymph/Mid/Gran (%/#)
Principle of measurementRBC/PLT/WBC/DIFF – electrical impedance; Hb – cyanide-free colorimetry; MCV measured, HCT calculated
Histograms3: WBC, RBC, PLT
Throughput60 tests per hour

Sample and reagents

ParameterH 360
Sample size9 μl (whole blood); 20 μl (predigested/capillary)
CalibratorELite H Cal, 3 ml, stability after opening 7 days at 2-8 °C
Control materialsErba H3 CON L/N/H, 3 ml, stability after opening 14 days at 2–8 °C

Data and constructiveness

ParameterH 360
Memory of results50,000 with charts
Interfaces4 USB + 1 LAN
Dimensions364×477×417 mm
Massa25kg
Power supplyAC 100–240 V, 50/60 Hz, ≤200 VA