Erba
ELite 3 hematology analyzer
Compact 3-part differential analyzer: 80 tests per hour and results in 45 seconds
ELite 3 is a compact 3-part differential hematology analyzer delivering 80 tests per hour for small or backup laboratory lines.
Request a quotationIt performs CBC with a 3-part white-cell differential, generates WBC/RBC/PLT histograms, and uses three cyanide-free reagents. A color touchscreen, built-in printer, USB, barcode-scanner connection, and LIS connectivity simplify daily operation.
22 parameters and 3 histograms; electrical impedance and LED photometry for hemoglobin. Throughput: 80 tests per hour, with results in up to 45 seconds. Sample volume: 25 µL whole or prediluted blood. Stores 10,000 results. Dimensions: 320 × 260 × 365 mm; weight: 12 kg.
Speed and simplicity
- Results in 45 seconds at 80 tests per hour
- Open sampling system with automatic sample rotor for safe loading
- Only 3 reagents (Erba Diluent-Diff, Erba Lyse-Diff, Erba Cleaner), without cyanide
- Full-color LCD touch screen, up to 13 interface languages
Measurement and quality control
- 22 parameters, full CBC with 3-component leukocyte formula
- Electrical impedance for cell count, LED-photometry for hemoglobin
- 3 histograms: WBC (3-diff), RBC, PLT
- 3 control blood levels, Levi-Jennings schedule, 2 types of calibration
Services and liaison
- 3 types of cleaning: high voltage pulse at the aperture, chemical cleaning, reverse pressure washing
- Technical Ruby Aperture for Durability
- Built-in printer with roll tape 58 mm, report with histograms
- USB port, connection to LIS
Measurement
| Parameter | ELite 3 |
|---|---|
| Parameters | 22: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, LYM, MID, GRA, LYM%, MID%, GRA%, RDW-SD, RDW-CV, PDW-SD, PDW-CV, MPV, PCT, P-LCC, P-LCR (research parameters) |
| Principle of measurement | electrical impedance (WBC/RBC/PLT); LED photometry for hemoglobin |
| Histograms | 3: WBC (3-diff), RBC, PLT |
| Throughput | 80 tests/h, result within 45 seconds |
Sample and reagents
| Parameter | ELite 3 |
|---|---|
| Sample size | 25 μl of whole blood or in preconception mode |
| Method of selection | open tube, automatic sample rotor |
| Reagents | Erba Diluent-Diff, Erba Lyse-Diff, Erba Cleaner (cyanide-free) |
Data and constructiveness
| Parameter | ELite 3 |
|---|---|
| Memory of results | 10,000, with 3-component RBC/PLT/WBC histograms |
| Interfaces | USB, LIS connection |
| Dimensions | 320 × 260 × 365 mm |
| Massa | 12 kg |
| Power supply | 12 In DC, 5 A, max. 60 W; external power supply 100–120/200–240 V AC, 50–60 Hz |