Micro Solutions

Analytical solutions for your laboratory

RU
Microaerophilic workstation M35

Don Whitley Scientific

Microaerophilic workstation M35

Microaerophile station with 560-620 cups for Campylobacter, Helicobacter pylori and other whimsical cultures

The M35 is a 560/620 cup microaerophile station for Campylobacter, Helicobacter and other demanding cultures.

Request a quotation

The system precisely controls oxygen, CO2, hydrogen, humidity and temperature, and allows samples to be manipulated in the same atmosphere where they are incubated. Touch screen, remote access, journaling and airlock 40 cups, 60s support the reproducible process.

Capacity: 560 / 620 cups; airlock 40 cups, 60 s. Gas connections: H2, N2, CO2, N2 and air according to configuration. Dimensions S×G×B 1255×720×710 mm, weight 100 kg; automatic drying is standard, additional moisture, cooling and inputs are selected as complete.

Atmosphere and flexibility

  • Up to four gases – nitrogen, carbon dioxide, air and a mixture of 10% hydrogen/90% nitrogen – are controlled in specified ratios for a particular experiment.
  • It can also work as an anaerobic station, providing flexibility in application.
  • Oxygen Profiling allows you to switch between microaerophilic and anaerobic conditions without changing the supplied gases

Ease of operation

  • Oval sleeve ports allow bare-handed work
  • Intuitive color touch screen with PIN access levels
  • Fast built-in airlock for 40 cups with a cycle of 60 seconds eliminates the risk of violation of conditions inside the chamber
  • Nitrogen purging of the lock reduces gas consumption

Automation

  • Automatic air drying without operator service
  • The Whitley Automatic Humidification System

Camera and transfer of samples

ParameterM35
Capacity560/620 cups 90mm
Airlock40 cups, 60 s cycle (increase option up to 5 minutes in 1 min increments)
PortsInstant Access or manual, oval sleeves

Gas system

ParameterM35
Gas supplyN2, CO2, air, mixture of 10% H2/90% N2 (up to 4 gases)
Sensors.Integrated O2, CO2 and H2 sensors

Control and automation

ParameterM35
Control screencolor-sensitive, PIN-access
Suckingautomatic, unmaintenance
MoisturizingThe Whitley Automatic Humidification System
HEPA filteringavailable (performance of HEPA)
Data loggingYes
Remote accessYes

Dimensions and mass

ParameterM35
Dimensions (W×D×H)1255×720×710 mm
Weight100 kg