Don Whitley Scientific
Hypoxic workstation H135 HEPA / GMP insulator
Hypoxia under HEPA filtering in a 900 liter chamber, including a validated GMP execution of Grade A
H135 HEPA/GMP is a 900-liter hypoxic workstation for cellular research in physiologically relevant atmospheres.
Request a quotationHypoxystation supports controlled O2, CO2, humidity and temperature, allowing cooking, incubating and examining cultures without returning to room air. Sensory control, O2 profiles, journaling, remote access and airlock 16 liters, cycle 60 c suitable for oncology, neuro- and cardiac studies. HEPA model exceeds ISO 14644 Class 3 requirements for the clean atmosphere inside the chamber. The designation in the source file combines HEPA and GMP: the GMP execution is a validated insulator with a Grade A/ISO 3 unidirectional stream.
chamber volume 900 liters; CO2 input, air and N2; airlock 16 liters, cycle 60 s. The family controls O2 up to 20% and CO2 up to 15% in increments of 0.1% in the respective configurations. Dimensions Sh×G×B 1473×1056×1063 mm, weight 175 kg; options HEPA, humidification, cooling and vacuum selection depend on execution.
HEPA performance (Hypoxystation)
- The highest, widest and deepest hypoxic chamber of the line, useful volume of about 900 liters
- It accommodates equipment for working inside the camera: visualization systems of living cells, microscopes, microplate readers
- Large removable front panel with two or three ports
- Whitley's built-in HEPA filtering system as standard
- Advanced Biological Containment (EBD)
- Two optional cooling options: remove heat from the equipment in the chamber or cool the atmosphere below 10 °C
- 16-liter nitrogen-blown lock to eliminate the risk of breach of chamber conditions
- Connection to the Whitley i2 Instrument Workstation
GMP performance (Grade A insulator)
- Precise control of O2 (0.1–20.0%) and CO2 (0.1–15.0%) for effective cultivation conditions
- HEPA H14 filter and unidirectional airflow support EU GMP Grade A
- Recycling technology creates ISO 3 environment – 100 times fewer particles than Grade A/ISO 5
- Continuous monitoring of viable and non-viable particles with microbiological air sampling
- Built-in VHP biodecontamination with isolated camera processing in adjacent modules
- Audit recording of all user inputs, events and alarms for data traceability
HEPA performance – camera and airlock
| Parameter | H135 HEPA / GMP |
|---|---|
| Chamber volume | 900 l. |
| Lock capacity | 16 l. |
| Lock cycle | 60. |
| Dimensions (W×D×H) | 1473×1056×1063 mm |
| Weight | 175 kg |
Atmospheric control (both performances)
| Parameter | H135 HEPA / GMP |
|---|---|
| Gas supply | CO2/air/N2 |
| Range O2 | 0.1-20.0%, step 0.1% |
| CO2 range | 0.1–15.0%, step 0.1% |
| Atmospheric purity class | exceeds ISO 14644 Class 3 (HEPA); ISO 3 in GMP |
GMP insulator - camera and airlock
| Parameter | H135 HEPA / GMP |
|---|---|
| Chamber volume | 900 l. |
| Lock capacity | 17.5 l or 43 l |
| Lock cycle | 1-5 minutes |
| Port system | 2 hand ports |
| Dimensions (W×D×H) | 1500×1130×1080 mm |
| Weight | 295 kg |