Don Whitley Scientific
Anaerobic workstation Whitley A135 HEPA / GMP
Large-sized anaerobic station with a volume of 900 liters for the placement of laboratory equipment
A135 HEPA/GMP is an anaerobic workstation with a useful capacity of 900 liters for cultivation and operation without contact with atmospheric oxygen.
Request a quotationThe station maintains a stable anaerobic atmosphere during inoculation, incubation and culture research. Color touch interface, automatic drying, remote access and monitoring conditions simplify documentation; airlock 40 cups; cycle 60 s. Built-in HEPA recycling cleans the atmosphere of the work chamber. Source line combines executions HEPA and GMP: The GMP option further relates to closed insulators with a unidirectional flow and validation packet.
Capacity: 900 liters; system of gases - anaerobic mixture and N2. airlock for 40 cups; cycle 60 s. Dimensions Sh×G×B: 1473×1056×1063 mm; weight 175 kg. The availability of Instant Access, moisture, cooling and other options depends on the selected configuration.
Capacity and access
- The camera capacity of 900 liters is enough to accommodate equipment like centrifuges and homogenizers.
- Removable front for cleaning and transfer of large-sized items and batches of samples
- Fast built-in airlock, 60 s cycle
- Ethernet for remote access to the touch screen
Filtration and Conditions (HEPA)
- Branded internal HEPA filter Whitley, purity is not worse than ISO 14644 class 3
- Fully integrated gas supply saves space on the table
- Automatic commissioning cycle and automatic drying
- Options: single sample input system, cooled incubation compartment, monitoring anaerobic conditions
GMP implementation
- Ceiling-mounted H14 HEPA filter; single-pass particle removal of 99.9997% (DOP)
- Laminar downstream supports EU GMP Grade A conditions in the work chamber
- Closed positive pressure recirculation system – the station can be used as an insulator in a Class D room
- Optional HEPA filtered airlock (H14) – reducing the particle content of the airlock in a 2-minute cycle
- IQOQ documentation is available, as well as PQ, URS, FAT, CD, RQ
Capacity and ports
| Parameter | A135 HEPA / GMP |
|---|---|
| Chamber volume | 900 l. |
| Airlock | 40 Petri cups (GMP performance: up to 10 bottles of Duran 500 ml), 17.5 liters |
| Lock cycle time | 60 s (GMP-execution - 2 minutes) |
Gas and filtration
| Parameter | A135 HEPA / GMP |
|---|---|
| Gas supply | anaerobic gas mixture / N2 (two gas lines) |
| HEPA filtering | internal, ISO 14644 Class 3; GMP execution – H14, 99.9997 % DOP, EU GMP Grade A |
| Removable front | There is (in GMP-execution - as standard) |
| Remote access / data registration | Yes |
Dimensions
| Parameter | A135 HEPA / GMP |
|---|---|
| Dimensions (W×D×H) | 1473 × 1056 × 1063 mm |
| Weight | 175 kg |