3
laser (405/488/532/642 nm)
Micro Solutions / Life Science
Desktop cytometer without sheath fluid: up to 14 parameters and accurate absolute counting of cells

Desktop cytometer without sheath fluid: up to 14 parameters and accurate absolute counting of cells
SL (tube) and HT (tube and 96-well plate) line on 1-3 lasers
A flexible line of compact flow cytometers for routine cellular analysis and high-performance research. Guava easyCyte combines up to three lasers, up to 14 parameters and direct absolute counting without counting particles.
Microcapillary hydrodynamics and precision syringe pump reduce sample consumption and do not require sheath fluid. The line includes single-tube SL and automated HT configurations for test tubes and 96-well microplates; InCyte supports multicolor analysis, heat maps and processing of the whole microplate. immunoprofiling, viability, apoptosis, cell cycle, small particle assessment and screening tasks.
Configurations on 1-3 lasers with wavelengths of 405, 488, 532 and/or 642 nm; from 5 to 14 channels, including up to 12 fluorescent parameters plus FSC and SSC. Particle range 0.2-60 μm, dynamic range 5 decades, recording speed 10-1000 events / s, flow rate 7-70 μl / min, CV fluorescence less than 4%. Sample formats: 0.5 and 1.5 ml, for HT also 96-well microplate; minimum volume 150, 900 and 100 μl, respectively. Dimensions SL 231 x 450 x 610 mm, weight 15.9 kg; HT 250 x 510 x 610 mm, weight 31.8 kg.
laser (405/488/532/642 nm)
parameters
particle size
counting
table-flow cytometer
8-color analysis on Spherotech beads
| Number of channels | 5–14 depending on the configuration |
|---|---|
| Lasers | up to 3:405, 488, 532 and/or 642 nm, according to configuration |
| Fluorescent parameters | 12 |
| Sample format | tubes 0.5 and 1.5 ml 0.5 and 1.5 ml tubes; 96-well plate |
|---|---|
| Automation of microplates and mixing | — Yes |
| Minimum sample size | 150 μl (0.5 ml); 900 μl (1.5 ml) 150 μl (0.5 ml); 900 μl (1.5 ml); 100 μl (microplate) |
| Range of detectable particles | 0.2-60 μm |
|---|---|
| Dynamic range | 5 decades |
| Speed of registration | 10–1000 events/s |
| Sampling consumption | 7-70 μl/min |
| Fluorescence accuracy | CV < 4% (CEN) |
| Dimensions (W×D×H) | 231×450×610 mm 250×510×610 mm |
|---|---|
| Weight | 15.9 kg 31.8 kg |
| Operating temperature | 16–30 °C, humidity up to 80% |
| Expenditure | free of sheath fluid; waste ≤50 ml in 8 hours of continuous operation |
| Software | GuavaSoft 4.0 with InCyte |
|---|---|
| Computer | Dell laptop running Windows |
| Guarantee | 1 year |
Identification of living, apoptotic and dead cells
Analysis of apoptosis (annexin V)
Cell cycle analysis
Checking and quality control of the device
Washing fluid for system maintenance
Waste and waste collection vials (separate for SL and HT)