Cytek
Spectral Cell Sorter Cytek Aurora CS
Spectral sorting of rare populations with direct transfer of panels from the Aurora analyzer
A highly parametric spectral variety to isolate rare cell populations and follow-up functional studies.Aurora CS combines Aurora optics, panels of up to 40 colors and flexible sorting of up to six populations.
Request a quotationThe common optical architecture with Aurora makes it easy to move panels from analyzer to sorter. SpectroFlo CS supports live unmixing, sorting by source or separated data, autofluorescence extraction, automatic drop delay, blockage detection, side jet monitoring, index sorting and automatic reporting.
Up to 5 lasers and 64 fluorescent detectors; sample feed rate of 10–80 μl/min, registration of up to 25,000 events/s. Sorting up to 6 populations in test tubes of 1.5 or 5 ml, double-thread sorting of 15 ml, index sorting in 96- and 384-well microplates. Replaceable nozzles of 70 and 100 μm; declared purity of at least 95% and yield of at least 90%.
Transfer of panels with Aurora
- Shared optical architecture with Aurora analyzer allows panels to be moved without reconfiguration of detectors
- Live unmixing and sorting by both source and separated data
- Autofluorescence extraction simplifies work with complex autofluorescent samples
Flexibility of sorting
- Replacement nozzles 70, 85 and 100 μm, 130 μm - optional
- Sorting into test tubes 1.5, 5 and 15 ml, up to 6 streams simultaneously
- Index sorting in 96- and 384-well microplates: less than 2 minutes on 96 holes and less than 5 minutes on 384 holes in high-performance mode
Control and security
- Automatic drop delay, blockage detection and side jet monitoring with live camera
- Initial evacuation of aerosol with built-in HEPA filters, optionally - class II biosafety box type A
- Automatic generation of a sorting report
Lasers and optics
| Parameter | 3 lasers | 4 lasers | 5 lasers |
|---|---|---|---|
| Laser configuration | 3-laser | 4-Laser (with Yellow-Green) | 5-laser |
| Fluorescent Detectors (APDs) | 38 | 48 | 64 |
Sorting
| Parameter | 3 lasers | 4 lasers | 5 lasers |
|---|---|---|---|
| Sample rate | 10–80 μl/min | ||
| Speed of registration | up to 25,000 events/s | ||
| Solo | 70, 85, 100 μm; 130 μm optionally | ||
| Number of sorted streams | up to 6 threads in tubes 1.5 ml; up to 4 threads in tubes 5 ml; double-thread sorting in 15 ml | ||
| Collection formats | 1.5/5/15 ml tubes; 96- and 384-well plates (index sorting) | ||
| Cleanliness and exit | purity at least 95%, yield at least 90% | ||
| Time to sort into a microplate | 96 holes - less than 2 minutes; 384 holes - less than 5 minutes (high-performance mode, 1 cell per hole) |
Safety and Workstation
| Parameter | 3 lasers | 4 lasers | 5 lasers |
|---|---|---|---|
| Biosecurity | primary evacuation of aerosol with HEPA filters; optionally, biosafety box type II A | ||
| Software | SpectroFlo CS: sorting control, automatic drop delay, blockage detection, index sorting, live sidestream camera |