Bruker
TIMS-QTOF mass spectrometer timsMetabo
Desktop 4D metabolomics and lipidomics: separating isomers by CCS and MS/MS to 300 Hz
TIMS-QTOF Desktop Platform for Highly Sensitive 4D Metabolomics and Lipidomics
Request a quotationtimsMetabo combines TIMS-MX, Athena Ion Processor and VIP-HESI to separate isomers and isobars, obtain purified MS/MS spectra and confident annotation on m/z, retention time and CCS. One run simultaneously supports quantitative determination and structural characterization.
The system uses MoRE mode to extend the range of ion mobility, high-frequency electronics TIMS-MX and precision detection. In the official example for creatinine, the range is 22-22624 ng / ml; values depend on the method and matrix.
Selectivity
- TIMS-MX Analyzer and MoRE Mode Expand M/Z Ranges and Mobility for Small Molecules
- Separation of isomers, isobars and interfering ions reduces chimeric MS/MS spectra
- Accurate and reproducible CCS values increase the confidence of the abstract
Sensitivity
- Athena Ion Processor boosts signal in MS and MS/MS modes
- VIP-HESI source for analytical flow rates; CSI for micro and nanoflow
- Creatinine is quantified in the range of 22–22,624 ng/ml
Quality control
- RSD area <5% and CCS deviation with standard deviation of 0.5% on 200+ urine injections
- QSee Test with Automatic “Yes/Not Good” Report
- TwinScape and TASQ RealTimeQC cloud monitoring software
Analyzer
| Parameter | timsMetabo |
|---|---|
| Type | TIMS-QTOF, desktop version |
| Ion mobility | Two-stage funnel TIMS-MX, More |
| quadrupole | High-precision, with Athena Ion Processor |
| MS/MS speed | up to 300 Hz (targeted quantitative analysis) |
| Collection regimes | PASEF, dia-PASEF |
Ionization and chromatography
| Parameter | timsMetabo |
|---|---|
| Sources | VIP-HESI, CaptiveSpray (CSI) |
| Chromatography | Analytical, Micro- and Nanoflow |
Indicators (producer data)
| Parameter | timsMetabo |
|---|---|
| Creatinine, Quantitative Definition Range | 22–22,624 ng/ml |
| Area reproducibility (NIST SRM 3672 urine) | RSD < 5%, more than 200 injections |
| Stability of CCS | Standard deviation ΔCCS 0.5% |
P.O. and status
| Parameter | timsMetabo |
|---|---|
| Processing | MetaboScape (CCS-Predict Pro, MetFrag, BioTransformer), TASQ |
| Purpose | RUO – for research purposes only |