Bruker
TargetScreener HR Screening Solution
Targeted and non-targeted pollutant screening based on 3,000+ compounds – from sample to report in 5 steps
A single HRMS-process for targeted and non-targeted screening of pollutants, toxicants and prohibited substances.
Request a quotationTargetScreener HR takes full-scan MS and bbCID data, maps it to the exact mass base, and generates a report in TASQ. The full set of signals remains available for retrospective search; new connections can be added to the database without reanalyzing the samples already measured.
The official database contains more than 3,000 records: monoisotope precursor masses, retention times, up to ten qualifier ions, bbCID fragments, adducts, isomers and CCS. The MRSQ/MRSQC score combines mass accuracy, retention time, isotope profile, diagnostic ions and, for 4D configuration, CCS.
All ions, all the time.
- Full MS and bbCID scan without loss of sensitivity, resolution or mass accuracy
- Retrospective search of previously measured samples after adding new compounds
- Diagnostic ions exclude false positive results and distinguish between colluting isobars
Prepared methods
- Ready-made LC/GC-, MS- and software methods reduce implementation time
- The criteria are in line with the SANTE 11312/2021 guidelines.
- 294 pesticides in wheat in a 5-minute method
4D screening
- TargetScreener 4D adds TIMS separation and CCS values as an additional criterion
- Ready-made GC-APCI method for dioxins, furans and PCBs according to Regulation (EU) 2017/644
- Method for 350 pollutants
Database
| Parameter | TargetScreener HR | TargetScreener 4D |
|---|---|---|
| Volume. | over 3,000 records | |
| Contents of the record | Monoisotope mass of precursor, retention time, up to 10 qualifier ions (bbCID and in-source fragments), adducts, isomers, CCS | |
| Classes of compounds | Pesticides, Veterinary Drugs, Drugs and NPS, Pharmaceuticals, PFAS, Steroids, Toxins, Dioxins, Environmental Pollutants |
Criteria and processing
| Parameter | TargetScreener HR | TargetScreener 4D |
|---|---|---|
| Evaluation | MRSQ | MRSQC (with CCS) |
| Mass Accuracy (SANTE) | ≤ 5 ppm (or ≤ 1 mDa below m/z 200) | |
| Allowance of retention time | ± 0.1 min | |
| Software | TASQ; MetaboScape - option | |
| Conclusion | Reports, exports to LIMS |
Hardware part
| Parameter | TargetScreener HR | TargetScreener 4D |
|---|---|---|
| Mass spectrometer | QTOF compact or impact II VIP | timsTOF |
| Chromatography | UHPLC, GC | |
| Ion Motility (TIMS) | — | yes |