Micro Solutions Life ScienceRU

Micro Solutions / Life Science

TargetScreener HR Screening Solution

Targeted and non-targeted pollutant screening based on 3,000+ compounds – from sample to report in 5 steps

TargetScreener HR Screening Solution

Product overview

Targeted and non-targeted pollutant screening based on 3,000+ compounds – from sample to report in 5 steps

High-resolution QTOF workflow with TASQ; TargetScreener 4D version adds CCS

A single HRMS-process for targeted and non-targeted screening of pollutants, toxicants and prohibited substances.

TargetScreener HR takes full-scan MS and bbCID data, maps it to the exact mass base, and generates a report in TASQ. The full set of signals remains available for retrospective search; new connections can be added to the database without reanalyzing the samples already measured.

The official database contains more than 3,000 records: monoisotope precursor masses, retention times, up to ten qualifier ions, bbCID fragments, adducts, isomers and CCS. The MRSQ/MRSQC score combines mass accuracy, retention time, isotope profile, diagnostic ions and, for 4D configuration, CCS.

Key figures

>3 000

base

5 steps

before the TASQ report

294

pesticide in 5 minutes

CCS

4D

Applications

Models and configurations

Product details

TargetScreener HR

Equipment, software and base in one solution

MRSQC assessment

match-indication

All ions, all the time.

Prepared methods

4D screening

Database

Volume.over 3,000 records
Contents of the recordMonoisotope mass of precursor, retention time, up to 10 qualifier ions (bbCID and in-source fragments), adducts, isomers, CCS
Classes of compoundsPesticides, Veterinary Drugs, Drugs and NPS, Pharmaceuticals, PFAS, Steroids, Toxins, Dioxins, Environmental Pollutants

Criteria and processing

EvaluationMRSQ
MRSQC (with CCS)
Mass Accuracy (SANTE)≤ 5 ppm (or ≤ 1 mDa below m/z 200)
Allowance of retention time± 0.1 min
SoftwareTASQ; MetaboScape - option
ConclusionReports, exports to LIMS

Hardware part

Mass spectrometerQTOF compact or impact II VIP
timsTOF
ChromatographyUHPLC, GC
Ion Motility (TIMS)—
yes

Sources of ionization

GC-APCI II

Gas chromatograph coupling through heated transfer line

Software

MetaboScape

Statistical analysis and identification of unknowns

Related equipment types

Related products

Classification

Product images

Documents

Manufacturer sources