Micro Solutions

Analytical solutions for your laboratory

RU
Pesticide Screening Solution for Pesticide Screener 2.0

Bruker

Pesticide Screening Solution for Pesticide Screener 2.0

860 pesticides in one database: screening, validation and quantification in one run

Ready-made UHR-QTOF workflow for simultaneous screening, validation and quantification of pesticides.

Request a quotation

PesticideScreener 2.0 combines UHPLC-QTOF, full-scan and bbCID data, pesticide database and TASQ processing. The approach allows hundreds of compounds to be tested in one run, reduces the risk of false-positive results, and reanalyzes stored data as the database expands.

The official validation poster lists a base of approximately 860 pesticides and a range of m/z 30–1000. For a representative set of 272 pesticides, the criteria for mass accuracy of at least 5 ppm or 1 mDa below m/z 200, the deviation of retention time not exceeding 0.1 minutes and the ratio of qualifier ions not exceeding 30% were applied.

One run is the whole list.

  • Alternating the full scan and bbCID fragmentation gives data on all ions of the sample
  • A database of approximately 860 pesticides and treatment in TASQ
  • The saved data is re-checked when the database is expanded

Trustworthy identification

  • Criteria: mass accuracy less than 5 ppm (less than 1 mDa below m/z 200), deviation of retention time not more than 0.1 min
  • Qualifier-ion ratio – an indicative criterion of not more than 30%
  • Calibration by clusters of sodium formate in each run and internal lock mass

Confirmed results

  • In independent validation, most of the 272 pesticides met EU SANTE criteria.
  • Validated LOQ of 10 μg/kg for most analytes
  • Linear calibrations 1–50 ng/mL (r2 > 0.98)

Method

ParameterPesticideScreener 2.0 on impact II
Collection regimeFull scan + bbCID, ESI+
Mass rangem/z 30–1 000
Database≈860 pesticides
processingTASQ

Identification criteria

ParameterPesticideScreener 2.0 on impact II
Mass accuracy< 5 ppm (< 1 mDa for m/z < 200)
Retention timeΔ < 0.1 min
Ion ratiosΔ ≤ 30% (approximately)

Validation results (NVWA, salad and orange)

ParameterPesticideScreener 2.0 on impact II
Number of pesticides272
Compliance with SANTE criteria at 10 μg/kg93% (salad), 78% (orange)
LOQ method10 μg/kg for most analytes
Linearity1–50 ng/ml, r2 > 0.98
Resolution of the instrumentup to 50,000 (m/z 800)