Bruker
MALDI-TOF Biotyper Sirius
Microorganism identification in minutes: up to 600 samples per hour
High-performance MALDI-TOF system for rapid identification of microorganisms in clinical microbiology.
Request a quotationThe system maps the mass-spectral profile of the isolate to the MALDI Biotyper library. IDealTune's automatic setup, long-lasting smartbeam laser and MBT Compass software environment support a standardized flow from target to result.
The manufacturer claims performance up to 600 samples per hour and the weight of the device is 75 kg. the Sirius version supports measurements in positive and negative ion modes; negative mode is intended for research applications.
Speed
- Results appear on the screen as the spectra are shot, without waiting for the end of the run.
- The full target (95 isolates and control) is processed in about 5 minutes.
- Urgent samples can be prioritized and answered in minutes.
- Smart Spectra Acquisition reduces the number of laser shots per sample
Opportunities
- Two ionic modes: positive for routine identification, negative for research tasks, such as lipid analysis
- The library covers Gram-positive and Gram-negative bacteria, yeast and mycelial fungi
- Modules for hemocultures, mycobacteria, fungi and markers of resistance
Ease of operation
- IDealTune auto-tuning performed in the background when analyzing the test standard
- Cleaning of the ion source is started from the software without depressurization
- Results in a clear “traffic light” color scheme
- Smartbeam laser with a life of 7 years or 500 million shots
Throughput
| Parameter | MALDI Biotyper sirius |
|---|---|
| Throughput | up to 600 samples/h |
| Full target report (95 isolates + 1 QC) | ~5 minutes |
| Sample preparation, manual time | ~20 s per isolate; 95 isolates less than 20 min |
| Direct identification from positive hemoculture | 15-20 minutes after the signal (Rapid Sepsityper Workflow) |
Mass spectrometer
| Parameter | MALDI Biotyper sirius |
|---|---|
| Method | MALDI-TOF mass spectrometry, desktop system |
| Ion regimen | Positive; negative (for research purposes only, RUO) |
| laser | Smartbeam, 200 Hz, resource 7 years or 500 million shots |
| Setting up | Automatic IDealTune with IVD Bacterial Test Standard |
| Cleaning up the ion source | built-in, IR laser, without vacuum disturbance |
| Weight | 75 kg |
Library and software
| Parameter | MALDI Biotyper sirius |
|---|---|
| Software | MBT Compass HT IVD with MBT IVD Library |
| Library coverage | More than 4,600 species/groups of bacteria, yeasts and mycelial fungi |
| Targets | MBT Biotarget 96 IVD, 96 positions, barcode |
| Integration | LIS/LIS, Sensitivity Determination Systems (AST), Laboratory Automation Systems |