The portfolio covers sequential and concurrent PET/CT, PET/MR and PET/SPECT/CT studies. The common positioning platform and ParaVision 360 support image recording and fusion, dynamic protocols, animal monitoring and quantitative analysis.

For PET/CT Si78, PET resolution of up to 0.7 mm, sensitivity of up to 12% and field of view of 150×80 mm is declared; CT provides a resolution of 52 μm and a dose of 6 mGy in low-dose mode. PET inserts support simultaneous operation with MRI 3, 7 and 9.4 T; parameters of other hybrid configurations depend on the modules.

Multimodality

  • PET/CT Si78 combines full-field SiPM-based PET imaging with fast low-dose SkyScan micro-CT
  • PET insert performs simultaneous (rather than sequential) imaging of PET and MRI on 3, 7 and 9.4 T systems
  • Albira Si trimodal line allows you to choose configurations of PET, SPECT, PET/CT, SPECT/CT, PET/SPECT and PET/SPECT/CT
  • ParaVision 360’s shared software platform provides a single workflow for scanning and combining images

Performance of PET/CT Si78

  • Full body vision up to 80×200 mm for mice and rats
  • Full CT scan of the body in 7.2 seconds at a resolution better than 52 microns
  • Low-dose CT regimen - less than 6 mGy
  • Gated PET and CT scan for cardiac and respiratory studies

Working with animals

  • Bilateral fully open access to gentry and minimum distance to the field of view of PET for kinetic studies of tracers
  • Single-clique self-connected crib for mice and rats with anesthesia, heating, breathing sensors and ECG

PET imaging (PET/CT Si78)

ParameterPET/CT Si78PET insertion (PET/MR)Albira Si (PET/SPECT/CT)
Field of vision (transaxial)80 mm——
Field of view (axial, one scan / max.)150/200mm——
Spatial resolution0.7 mm——
Sensitivity12%——

CT imaging (PET/CT Si78)

ParameterPET/CT Si78PET insertion (PET/MR)Albira Si (PET/SPECT/CT)
Field of vision (transaxial)80 mm——
Image resolutionbetter than 52 μm.——
Time for full body scan7.2 s——
Dose (low-dose regimen)<6 mG——

Multimodal configurations

ParameterPET/CT Si78PET insertion (PET/MR)Albira Si (PET/SPECT/CT)
Compatibility with MRI—3; 7; 9.4 Tl—
Available configurationsPET/CTPET/MRI (simultaneously)PET, SPECT, PET/CT, SPECT/CT, PET/SPECT, PET/SPECT/CT
G.O.O.ParaVision 360ParaVision 360—