Bruker
Optical in vivo imaging system Ami HTX
Bioluminescence, fluorescence and X-rays for five mice in one exposure
Desktop BLI/FLI/X-ray system for high-performance imaging of up to five small animals.
Request a quotationAmi HTX combines bioluminescence, fluorescence and X-rays in a compact system with automatic field of view and filter selection. Aura performs calibrated quantification, and a heated platform and anesthesiological connections support longitudinal studies.
The system is designed for 5 mice; the maximum optical field of view is 25×17 cm The camera is cooled to −90 °C, 10 LED-excitation wavelengths and 10 emission filters are used; the HTX configuration includes an X-ray channel.
Sensitivity and quality of data
- Camera with CCD sensor cools to -90 °C due to Peltier elements and air - no liquids, chillers and risk of leaks, ready for operation in about 5 minutes
- The patented LED excitation delivers significantly more light to the object than a white light bulb, without fiber optic – a higher signal and below the background
- The absolute calibration traceable to the NIST standard makes the data comparable between the series and the camera settings.
- Minimum recorded brightness - 50 photons / s / cm2 / s
Throughput
- 25x17 cm field of view allows you to shoot 5 mice at the same time
- Automatic X-ray channel: anatomical binding without animal relaying
- Easy Mode: Select probe and object and start shooting; Smart AutoExp selects exposure automatically
- Kinetics Mode Automatically Builds Bioluminescence Kinetic Curves in Real Time
Convenience and reliability
- Heated platform (20–40°C) and gas anesthesia ports are compatible with third-party systems
- Vertical door saves space and reduces the output of the anesthetic in the room
- Lightproof steel case without seals and end switches, factory calibration without installation on site
- Aura analysis software is free for Windows and Mac and opens Living Image files
Modality and capacity
| Parameter | Ami HTX |
|---|---|
| Visualization techniques | bioluminescence (BLI), fluorescence (FLI), X-ray |
| Capacity | up to 5 mice |
| Optical field of view | 10×7 to 25×17 cm |
| Visualization platform | 50x34 cm, heated 20-40 °C |
| Gas anesthesia | port-in |
Camera.
| Parameter | Ami HTX |
|---|---|
| Sensor | CCD 1" with backlight, Grade 1 plus |
| Sensor resolution | 1152 × 770 pixels |
| Sensor size | 25.9 x 17.3 mm |
| Cooling | -90 °C, solid state, air |
| Quantum efficiency | >85% in the 500–700 nm range; >35% in the 700–900 nm range |
| Dark current. | <49 e−/s/cm2 |
| binning | 1×1, 2×2, 4×4, 8×8 |
| Objective | 50 mm, aperture f/1.2–f/16 |
| Minimum recorded brightness | 50 photons/s/cm2/sp |
fluorescence
| Parameter | Ami HTX |
|---|---|
| LED-excitation | 10 wavelengths: 430, 465, 500, 535, 570, 605, 640, 675, 710, 745 nm |
| Emission filters | 10 filters to choose from 20 standard (490-870 nm, step 20 nm) |
| LED resource | 100,000 hrs |
X-ray
| Parameter | Ami HTX |
|---|---|
| Source | 10–40 keV |
| Detector | linear scanning |
| Field of vision | 25 x 15 cm |
General.
| Parameter | Ami HTX |
|---|---|
| Dimensions (E × G × B) | 56 × 66 × 122 cm |
| Calibration | absolute, traceable to NIST; factory |
| Set | PC, monitor, Aura software |
| Software | Aura (collection and analysis); analysis – unlicensed, Windows and Mac |