Biolog
Odin Automatic Identification and Phenotyping Platform
Automatic identification and phenotyping of microbes: 8 to 50 microplates at a time
Biolog Odin is an automated platform for microbial identification and functional phenotyping. Two configurations close different levels of flow: the compact Odin VIII works with 1-8 microplates at the same time, and the Odin L with 1-50 microplates.
Request a quotationThe system combines incubation, microplate kinetic reading and data analysis into a single workflow. It is suitable for microbial identification, growth curves, cellular respiration assessment, metabolic profiling, microbial community analysis and Phenotype MicroArray. Odin Software maps metabolic profiles. with the Biolog database and maintains user bases for own strains.
Odin VIII can hold 1 to 8, Odin L can hold 1 to 50 microplates. Both systems support incubation at 22 to 45 °C and final or kinetic reading at intervals of 2 to 20 minutes. The optical circuit uses 490 or 590 nm to record redox paint and 740 nm to assess turbidity/growth. The platform works with microbial identification microplates. Phenotype MicroArray and community analysis solutions; the software environment is declared by the manufacturer to be compatible with the requirements of 21 CFR Part 11. The specific set of microplates and the validation package should be confirmed for the selected application.
Productivity and flexibility
- Odin VIII works with 1-8 microplates, Odin L – with 1-50 microplates at the same time
- Kinetic reading at intervals of 2-20 minutes or single measurement at the end point
- Each microplate is measured at the right temperature and at the right time without the operator’s involvement.
Analysis opportunities
- Identification of bacteria, anaerobes, yeasts and mycelial fungi on a single database of more than 2,900 organisms
- Optical density at 490 or 590 nm for redox reaction and 740 nm for growth assessment
- Automatically stop the experiment when matching the database
Compliance and integration
- 21 CFR Part 11 for regulatory purposes
- Analysis protocols and tools are built into software
- Data output to an external computer
Throughput
| Parameter | Odin VIII | Odin L |
|---|---|---|
| Capacity | 8 microplates | 50 microplates |
| Reading mode | once (endpoint) or kinetically with an interval of 2-20 minutes | |
| Dimensions (W×D×H) | 61.5×53.3×45.6 cm | 61.5×53.3×84.2 cm |
Optics and temperature
| Parameter | Odin VIII | Odin L |
|---|---|---|
| OD wavelengths | 490 or 590 nm and 740 nm | |
| Incubation range | 22–45°C | |
| Temperature stability | ±2°C in the tray chamber | |
| Working temperature of the environment | 18–28°C |
Power supply and interface
| Parameter | Odin VIII | Odin L |
|---|---|---|
| Power supply | 100–240 V, 50/60 Hz | |
| Monitor | 22-inch LCD display | |
| Regulatory options | 21 CFR Part 11 (optional) |