Micro Solutions

Analytical solutions for your laboratory

RU
SafeMate Total's biosafety cabinet

BioAir

SafeMate Total's biosafety cabinet

Class II B2 cabinet with 100% air removal for bioagents and small volumes of chemistry

SafeMate Total is a BioAir II class B2 biosafety cabinet with complete air removal, designed for processes that simultaneously require biological protection and work with small amounts of chemicals without vapor accumulation due to recycling.

Request a quotation

The cabinet takes 100% fresh air from the laboratory: the main stream is cleaned by the inflow HEPA H14 and forms the ISO 3 work zone, then all the air passes through the exhaust HEPA H14 and is removed by the external duct. The microprocessor controls the laminar and inlet flows, the position of the glass and the resource of the filters and UV lamps; in case of insufficient flow The electric front glass, the V-grid and the camera from AISI 304 make it easy to operate safely and clean. Connecting to an external exhaust system is mandatory.

Executions: 1.2 and 1.8. Power - 220-240 V, 50/60 Hz; power of the fan and lighting - 600 and 1080 W. Dimensions without support - 1380 × 795 × 1640 and 1990 × 795 × 1640 mm; working area - 1230 × 580 × 700 and 1840 × 580 × 700 mm; weight - 280 and 400 kg. Laminar flow speed - 0.35-0.38 m / s, entrance barrier - 0.54 m / s ± 10%; required consumption of external hood - 1300 and 2270 m3/h ± 10% at a dilution from −400 to −500 Pa. ISO 3 working area; illumination >750 lux, noise <65 dB(A), vibration <0.005 mm RMS. Supply and exhaust filters H14, MPPS efficiency - 99.995%; exhaust pipe diameter - 315 mm. Compliance EN 12469:2000 and NSF/ANSI 49-2008 specified by the manufacturer.

Protection in working with chemistry

  • 100% fresh air from the laboratory passes through the inflow HEPA H14 and forms the ISO 3 work area.
  • All air is removed through the HEPA H14 exhaust and the external duct without recirculation into the room.
  • The design allows you to safely work with small amounts of chemicals without vapor accumulation
  • Mandatory connection to the calculated external exhaust system (hard connection by air duct)

Monitoring and alarm

  • The microprocessor continuously monitors the laminar and input flow rates with anemometers.
  • Optical and sound alarms are triggered when there is insufficient flow
  • Displays the workout of the cabinet, HEPA filters and UV lamps
  • Non-potential contact for remote control of the fan or connection to the alarm system

Construction and maintenance

  • Electric front glass with automatic recognition of correct position
  • V-shaped grille removes the need for armrests
  • Camera and work surface from AISI 304, access to filters and electronics from the front
  • DOP port to check the integrity of HEPA filters

Throughput

Parameter1.21.8
Class and type of protectionClass II, Type B2 (full hood)
Laminar flow rate0.35–0.38 m/s
Speed of the air barrier entrance0.54 m/s ± 10%
Required external hood consumption1300 m3/h ± 10%2270 m3/h ± 10%
Dilution in the extraction system-400 to -500 Pa
Class of cleanliness of the working areaISO 3

Filtration

Parameter1.21.8
Filterssupply and exhaust HEPA H14
Effectiveness of MPPS99,995%
Diameter of exhaust pipe315 mm
Maximum height of the front opening440 mm

Dimensions and conditions of use

Parameter1.21.8
Dimensions without support (D×G×B)1380 × 795 × 1640 mm1990 × 795×1640 mm
Working area (DxGxV)1230×580×700 mm1840×580×700 mm
Weight280 kg400 kg
Lighting>750 lux
Sound level<65 dB(A)
Vibration.<0.005 mm RMS
Power supply220–240 V, 50/60 Hz
Power consumption (fan and light)600 watts1080 W.