BioAir
SafeMate EZ Biological Safety Room
Class II type A2 with electric glass drive and optional gas and vacuum cranes
The SafeMate EZ is an A2-class BioAir II biosafety cabinet for everyday work with biomaterials and cell cultures. The front-glass electric drive, a fully stainless work chamber, and an ergonomic V-shaped input grille combine operator, product, and environmental protection with the convenience of rou
Request a quotationThe cabinet forms a vertical laminar flow and a controlled air barrier, and HEPA filtering reduces the risk of cross-contamination. The flow rates are displayed in real time; the system monitors the correct position of the front window. The 5° tilted work area facilitates access, the electric glass makes it easier to open and the front window is not easily accessible. The steel chamber and work surface are designed for regular cleaning, the design is compatible with VHP decontamination, and the UV-C lamp is standard.
Performances: 1.2 and 1.8. Power - 220-240 V, 50/60 Hz; power consumption of the fan and lighting - 465 and 774 W. Dimensions without support - 1380×795×1450 and 1990×795×1450 mm; working area - 1230×600×700 and 1840×600×700 mm; weight - 256 and 360 kg. Filters H14 EN 1822-1, global efficiency MPPS - 99.995. Sound level - <56 and <60 dB(A), vibration <0.0005 mm RMS. Two auxiliary outlets; gas, air and vacuum service valves are available as options. EN 12469 is declared by the manufacturer.
Protection and control
- Vertical laminar flow and controlled inlet air barrier with real-time velocity display
- The system controls the correct position of the electric front glass
- The design is compatible with VHP-decontamination with hydrogen peroxide vapors
- UV-C lamp on the back wall is standard
Ease of operation
- V-shaped input grille eliminates the need for armrests
- 5° tilted work area facilitates access to samples
- The electric glass opens and closes with buttons on the control panel
- Fully stainless working chamber and surface are designed for regular cleaning
Flexible configuration
- Gas and vacuum cranes can be installed as an option, including after delivery.
- Active and passive extraction kits with additional carbon filter are available
- A stand with drawers on wheels, a support rack and a set of wheels - for the tasks of the laboratory
Throughput
| Parameter | 1.2 | 1.8 |
|---|---|---|
| Class and type of protection | Class II, type A2 | |
| Laminar flow rate | 0.33–0.40 m/s | |
| Speed of the air barrier entrance | 0.53 m/s ± 10% | |
| Exhaust air flow | 480 m3/h ± 10% | 600 m3/h ± 10% |
| Class of cleanliness of the working area | ISO 5 according to EN 14644-1 |
Filtration and protection
| Parameter | 1.2 | 1.8 |
|---|---|---|
| Class of filter efficiency | H14 according to EN 1822-1 | |
| Global efficiency of MPPS | 99,995% | |
| Frontal aperture protection coefficient (Apf) | ≥1,0×10⁵ | |
| Passage of UV-C through glass | 98% |
Dimensions and conditions of use
| Parameter | 1.2 | 1.8 |
|---|---|---|
| Dimensions without support (D×G×B) | 1380 × 795 x 1450 mm | 1990 × 795×1450 mm |
| Working area (DxGxV) | 1230×600×700 mm | 1840×600×700 mm |
| Weight | 256 kg | 360 kg |
| Sound level | <56 dB(A) | <60 dB(A) |
| Vibration. | <0.005 mm RMS | |
| Power supply | 220–240 V, 50/60 Hz | |
| Power consumption (fan and light) | 465 W | 774 W. |