BioAir
SAFE3 Class III Biological Safety Insulator
Class III sealed insulator with glove ports for work with pathogens of risk group 4
Safe3 is BioAir Class III airtight biosafety cabinet for maximum retention of dangerous biological agents. Glove ports, airlocks with mutual door locking and complete air removal allow you to organize a closed process to work with pathogens up to risk group 4.
Request a quotationThe operator manipulates through two glove ports without opening the sealed work chamber. Materials are transmitted through a side door lock, the inflow passes through the prefilter and HEPA H14, and all air removed is sequentially through two exhaust HEPA H14. The microprocessor monitors the flow and state of SAFE/UNSAFE, performs Self-calibration and high-speed purges at launch, compensates for the gradual contamination of filters and emits multi-level light and sound alarms. The Magnehelic manometer provides constant visual monitoring of the pressure in the chamber.
Microbiological protection class - III according to EN 12469:2000. Power - 220/230 V, 50/60 Hz; maximum power - 1500 W. Dimensions - 2015×1295×810 mm, working chamber - 1200×675×695 mm, weight - 220 kg; two glove ports. Full air consumption - 200 m3 / h ±10%. Clean working area - ISO 3; illumination >900 lc, noise <56 dB(A), vibration <0.005 mm RMS. The filtration system includes an input prefilter H10, an input HEPA H14 and two consecutive HEPA H14 exhausts; the efficiency of H14 by MPPS is 99.995%. The camera is made of AISI 304, the front window is a multilayer safety glass 4 + 4 mm.
Leakproofness and retention
- The operator works through two glove ports without opening the sealed work chamber.
- Materials are transmitted through a side airlock with mutual door locking
- Exclusive four-stage filtering: Inlet prefilter, input HEPA H14 and two consecutive HEPA H14 exhausts
- Magnehelic manometer provides permanent visual control of pressure in the chamber
Control and security
- The microprocessor performs self-calibration and high-speed purges when starting before switching to SAFE mode.
- Three modes of operation: working, waiting, calibration
- Multilevel light and sound alarms with redundancy function, including power failure signal
- Main switch with removable key – lock from accidental switching on in SAFE mode
Ergonomics and maintenance
- C-shaped support rack for installation by one person
- Filter replacement and maintenance - from the front side of the cabinet
- Backflow protection valve is available as an option for channel configuration
Protection and leakage
| Parameter | SAFE³ |
|---|---|
| Class of microbiological protection | III according to EN 12469:2000 |
| Total airflow | 200 m3/h ± 10% |
| Class of cleanliness of the working area | ISO 3 |
| Number of glove ports | 2 |
Filtration
| Parameter | SAFE³ |
|---|---|
| Filtration system | H10 input prefilter, H14 input HEPA, two consecutive HEPA H14 exhausts |
| Efficiency of HEPA H14 by MPPS | 99,995% |
Dimensions and conditions of use
| Parameter | SAFE³ |
|---|---|
| Dimensions (W×D×H) | 2015×1295×810 mm |
| Working chamber (SxG×V) | 1200×675×695 mm |
| Weight | 220 kg |
| Lighting | >900 lux |
| Sound level | <56 dB(A) |
| Vibration. | <0.005 mm RMS |
| Camera material | AISI 304 |
| Front window | multilayer safety glass 4+4 mm |
| Power supply | 220/230 V, 50/60 Hz |
| Maximum power consumption | 1500 W. |