5 at 1
function
Micro Solutions / Life Science
5 devices in one: dispenser, shaker, incubator, washer and reader for one 96-well microplate

5 devices in one: dispenser, shaker, incubator, washer and reader for one 96-well microplate
fully automated cycle of the ELISA without transferring the microplate between devices
Crocodile LB 925 miniWorkstation 5-in-one is a compact system for complete automation of one 96-well ELISA microplate, which frees the operator from sequential transfer of the microplate between individual devices.
It combines a dispenser, shaker, incubator, washer and photometric reader. Open source software allows you to create protocols for various 96-well tests, manage incubation, dosing, washing, stirring and reading. Version 4-in-one has the same working mechanics, but without a built-in reader.
Format - one 96-well microplate ANSI / SLAS. Reader version 5-in-one: range 400-690 nm, pre-installed filters 405/450/492/620 nm, 8 slots, range 0-3.0 OD at 450 nm Dispenser: four independent pumps, 10-2000 μl. washer: autoclavable 8-channel manifold, 25-1000 μl, three containers of 1 liter and drain capacity of 2 l. ambient temperature +4 °C to 55 °C; linear stirring 5-20 Hz Dimensions 25.8×26.3×62.6 cm; weight 14.8 kg.
function
hull-width
loophole
research
body width 26.3 cm
flushing-quality
| Format of the sample | 1×96-well microplate (ANSI/SLAS) |
|---|---|
| Dimensions (W×SH×G) | 25.8×26.3×62.6 cm |
| Weight | 14.8 kg (version 5-in-1) |
| Interface. | USB |
| Power supply | 100–240 V AC power supply 24 V DC/3.3 A |
| Operating temperature | 10–35 °C |
| Humidity | up to 75% without condensation |
| Temperature range | from ambient temperature +4 °C to 55 °C |
|---|---|
| Temperature stability | ±1 °C on microplate at 37 °C |
| Incubation time. | programmable |
| Dynamic range | 0-3.0 OD at 450 nm |
|---|---|
| Spectral range | 400–690 nm (filters 405/450/492/620 nm) |
| Filter slots | 8 |
| Accuracy/precision | ±0.01 OD or 2.5% (which is more); <1.5% CV (0.01–0.5 OD and 2.0–3.0 OD), <1% CV (0.5–2.0 OD) |
| Dispenser | 4 independent pumps, 10–2000 μl, accuracy <5% and precision <2% CV at 100 μl |
|---|---|
| flusher | autoclavable 8-channel manifold, 25-1000 μl, 3 containers of 1 liter and drain 2 l |
| Shaker. | linear stirring at 5–20 Hz |
Advanced Analysis and Reporting Module, FDA compliance with 21 CFR Part 11
Maximum configuration of LightCompass software
Additional license of the software manager
Package of documents for qualification of installation and operation
Test microplate for validation of 96-well absorption readers