Turns ASTEC’s portable camera into a satellite incubator right on a workstation
Heat-Wise RH-3100 is a compact heated platform for converting an ASTEC portable camera into a satellite incubator at an IVF station.
Request a quotationThe platform is compatible with the single-close HC-3100 and four-section HC-3100-2, as well as the existing CUBE AD-3100. removable aluminum units are autoclaved; the optional adapter delivers pre-mixed gas.
RH-3100: 252×270×77 mm, 3.2 kg; digital PID, range from ambient +7 °C to 40 °C, oscillation ±0.3 °C, uniformity of the aluminum surface ±0.5 °C. Gas ports IN/OUT Ø6 mm; power AC 100–240 V. HC-3100 holds up to 12 cups of 35 mm.
Ease of use
- It is enough to install a portable camera on the platform – it works as a satellite incubator of a workstation.
- Compatible with the single-close camera HC-3100 and the separate four-section HC-3100-2, as well as with existing models AD-3100 without modification
Temperature monitoring
- Digital PID regulation maintains temperature from +7 to 40 °C
- Uniformity of aluminum surface ± 0.5 °C
- Optional adapter RH-GP feeds pre-prepared gas mixture
Platform RH-3100
| Parameter | RH-3100 |
|---|---|
| Sizes. | 252×270×77 mm |
| Temperature management | digital IDP |
| Temperature range | from ambient temperature +7 to 40 °C |
| Temperature fluctuation | ±0,3 °C |
| Surface uniformity | ± 0.5 °C (aluminum surface) |
| Gas ports | input/exit, Ø6 mm (outer diameter) |
| Weight | 3.2 kg |
| Power supply | AC 100–240 V, 50/60 Hz, up to 1 A |
Compatible portable chambers
| Parameter | RH-3100 |
|---|---|
| HC-3100, single-cap | 180×217×20 mm, 781 ml, 4-well format – 4 dishes |
| HC-3100-2, 4 sections (one section) | 94×100×20 mm, 186 ml; dishes Ø35 mm - 4 pcs., or Ø60 mm - 1 pc., or 4-well format - 1 pc. |