Analytik Jena
ISP-MS spectrometer PlasmaQuant MS Elite S
Ultrasensitive ISP-MS between routine MS Q and research MS Elite
PlasmaQuant MS Elite S is a highly sensitive ISP-MS for ultra-trace analysis, combining enhanced ion transfer with the flexibility of routine and research techniques.
Request a quotationThe model is intermediate between routine MS Q and highly sensitive MS Elite. Optimized ion optics provides high signal at low background, which is important for ultratrace concentrations, rare elements and limited volume samples. The platform supports common sample input kits, iCRC and software environments. The series allows transferring methods between configurations.
Declared sensitivity for 115In - 1.1 million IMP / s per μg / l, background at a mass of 5 AU - less than 0.7 IMP / s. General characteristics of the platform: four-channel peristaltic pump 0-100 rpm, concentric sprayer 400 μl / min, two-way chamber Scott with Peltier cooling up to -15 ° C, low-flow burner and 90-degree reflective ion Optics: Inert PFA kits and organic/volatile organic sampling are available.
Ion optics and separation
- ReflexION’s 90-degree reflective ion optics with three-dimensional ion beam focus separates sensitive components from a direct beam of photons and neutrals
- HD-quadrupole with a full mass range of 3-260 AU provides high resolution and low abundance sensitivity
- The all-digital AD Detection does not require regular cross-calibration, as is the case with analog-to-digital detectors.
Elimination of interference and stability
- The built-in helium and hydrogen iCRC reaction cell provides rapid and efficient suppression of molecular interferences.
- BOOST technology supports sensitivity in gas mode without degrading detection limits
- Interface cones are replaced in less than a minute without disassembly of the sample input system, which reduces downtime
Sampling and software
- The open layout of the input system simplifies access to the sprayer, camera and burner during everyday maintenance.
- ASpect MS software automates calibration and resolution tests
- Modules for aerosol dilution, work with pure organic solvents (Nitrox) and pairing with HPLC are available
Intermediate configuration
- It occupies an intermediate position between routine MS Q and the most sensitive MS Elite in terms of sensitivity.
- Optimized ion optics provides high signal at low background for ultratrace concentrations
- Common sample entry kits, iCRC and series software environment allow transfer of techniques between platform configurations
Sensitivity and specificity of the model
| Parameter | PlasmaQuant MS Elite S |
|---|---|
| Sensitivity (115In) | 1 100,000 mp/s per μg/l |
| Background (mass 5 AU) | less than 0.7 mp/s |
| Purpose | Ultratratraceability Analysis between Routine and Research Configurations |
| Input kits | inert PFA kits, the introduction of organic / volatile organic samples |
Ion optics and detector
| Parameter | PlasmaQuant MS Elite S |
|---|---|
| Ion optics | ReflexION, 90° reflection, 3D focusing of the ion beam |
| quadrupole | HD-quadrupole 3 MHz, mass range 3-260 AU, min. fixation time 50 μs, scanning speed > 5000 AU / s |
| RF generator | solid-state, 27 MHz, plasma power 300-1600 W |
| Detector | All-digital AD Detection system |
Elimination of interference
| Parameter | PlasmaQuant MS Elite S |
|---|---|
| Reaction cell | iCRC (Integrated Collision Reaction Cell), He and H2 gases at the top of the skimmer cone |
| BOOST technology | maintains sensitivity in gas mode without loss of detection limits |
| Replacement of interface cones | less than 1 min without dismantling the sampling system |
Sampling system
| Parameter | PlasmaQuant MS Elite S |
|---|---|
| Peristaltic pump | four-channel, 0-100 rpm |
| Sprayer | concentric, 400 μl/min |
| Spray chamber | Scott's two-way chamber, Peltier cooling from room temperature to -15 °C |
| Burner. | low-consumption, solid, 2.4 mm injector, support for hot and cold plasma |