ISP-MS spectrometer PlasmaQuant MS Elite S

Analytik Jena

ISP-MS spectrometer PlasmaQuant MS Elite S

Ultrasensitive ISP-MS between routine MS Q and research MS Elite

PlasmaQuant MS Elite S is a highly sensitive ISP-MS for ultra-trace analysis, combining enhanced ion transfer with the flexibility of routine and research techniques.

Request a quotation

The model is intermediate between routine MS Q and highly sensitive MS Elite. Optimized ion optics provides high signal at low background, which is important for ultratrace concentrations, rare elements and limited volume samples. The platform supports common sample input kits, iCRC and software environments. The series allows transferring methods between configurations.

Declared sensitivity for 115In - 1.1 million IMP / s per μg / l, background at a mass of 5 AU - less than 0.7 IMP / s. General characteristics of the platform: four-channel peristaltic pump 0-100 rpm, concentric sprayer 400 μl / min, two-way chamber Scott with Peltier cooling up to -15 ° C, low-flow burner and 90-degree reflective ion Optics: Inert PFA kits and organic/volatile organic sampling are available.

Ion optics and separation

  • ReflexION’s 90-degree reflective ion optics with three-dimensional ion beam focus separates sensitive components from a direct beam of photons and neutrals
  • HD-quadrupole with a full mass range of 3-260 AU provides high resolution and low abundance sensitivity
  • The all-digital AD Detection does not require regular cross-calibration, as is the case with analog-to-digital detectors.

Elimination of interference and stability

  • The built-in helium and hydrogen iCRC reaction cell provides rapid and efficient suppression of molecular interferences.
  • BOOST technology supports sensitivity in gas mode without degrading detection limits
  • Interface cones are replaced in less than a minute without disassembly of the sample input system, which reduces downtime

Sampling and software

  • The open layout of the input system simplifies access to the sprayer, camera and burner during everyday maintenance.
  • ASpect MS software automates calibration and resolution tests
  • Modules for aerosol dilution, work with pure organic solvents (Nitrox) and pairing with HPLC are available

Intermediate configuration

  • It occupies an intermediate position between routine MS Q and the most sensitive MS Elite in terms of sensitivity.
  • Optimized ion optics provides high signal at low background for ultratrace concentrations
  • Common sample entry kits, iCRC and series software environment allow transfer of techniques between platform configurations

Sensitivity and specificity of the model

ParameterPlasmaQuant MS Elite S
Sensitivity (115In)1 100,000 mp/s per μg/l
Background (mass 5 AU)less than 0.7 mp/s
PurposeUltratratraceability Analysis between Routine and Research Configurations
Input kitsinert PFA kits, the introduction of organic / volatile organic samples

Ion optics and detector

ParameterPlasmaQuant MS Elite S
Ion opticsReflexION, 90° reflection, 3D focusing of the ion beam
quadrupoleHD-quadrupole 3 MHz, mass range 3-260 AU, min. fixation time 50 μs, scanning speed > 5000 AU / s
RF generatorsolid-state, 27 MHz, plasma power 300-1600 W
DetectorAll-digital AD Detection system

Elimination of interference

ParameterPlasmaQuant MS Elite S
Reaction celliCRC (Integrated Collision Reaction Cell), He and H2 gases at the top of the skimmer cone
BOOST technologymaintains sensitivity in gas mode without loss of detection limits
Replacement of interface conesless than 1 min without dismantling the sampling system

Sampling system

ParameterPlasmaQuant MS Elite S
Peristaltic pumpfour-channel, 0-100 rpm
Sprayerconcentric, 400 μl/min
Spray chamberScott's two-way chamber, Peltier cooling from room temperature to -15 °C
Burner.low-consumption, solid, 2.4 mm injector, support for hot and cold plasma