ISP-MS spectrometer PlasmaQuant MS Elite S

Ultrasensitive ISP-MS between routine MS Q and research MS Elite

A general picture of the PlasmaQuant MS and MS Elite series, the execution and configuration depend on the model chosen.

Product overview

Ultrasensitive ISP-MS between routine MS Q and research MS Elite

for ultratrace concentrations, rare elements and limited volume samples

PlasmaQuant MS Elite S is a highly sensitive ISP-MS for ultra-trace analysis, combining enhanced ion transfer with the flexibility of routine and research techniques.

The model is intermediate between routine MS Q and highly sensitive MS Elite. Optimized ion optics provides high signal at low background, which is important for ultratrace concentrations, rare elements and limited volume samples. The platform supports common sample input kits, iCRC and software environments. The series allows transferring methods between configurations.

Declared sensitivity for 115In - 1.1 million IMP / s per μg / l, background at a mass of 5 AU - less than 0.7 IMP / s. General characteristics of the platform: four-channel peristaltic pump 0-100 rpm, concentric sprayer 400 μl / min, two-way chamber Scott with Peltier cooling up to -15 ° C, low-flow burner and 90-degree reflective ion Optics: Inert PFA kits and organic/volatile organic sampling are available.

Key figures

1 100 000

μg/L (115In)

iCRC

elimination

−15 °C

spray-cooling

Applications

Models and configurations

Product details

PlasmaQuant MS Elite S

intermediate configuration

PlasmaQuant MS lineup

four-platform models

Ion optics and separation

Elimination of interference and stability

Sampling and software

Intermediate configuration

Sensitivity and specificity of the model

Sensitivity (115In)1 100,000 mp/s per μg/l
Background (mass 5 AU)less than 0.7 mp/s
PurposeUltratratraceability Analysis between Routine and Research Configurations
Input kitsinert PFA kits, the introduction of organic / volatile organic samples

Ion optics and detector

Ion opticsReflexION, 90° reflection, 3D focusing of the ion beam
quadrupoleHD-quadrupole 3 MHz, mass range 3-260 AU, min. fixation time 50 μs, scanning speed > 5000 AU / s
RF generatorsolid-state, 27 MHz, plasma power 300-1600 W
DetectorAll-digital AD Detection system

Elimination of interference

Reaction celliCRC (Integrated Collision Reaction Cell), He and H2 gases at the top of the skimmer cone
BOOST technologymaintains sensitivity in gas mode without loss of detection limits
Replacement of interface conesless than 1 min without dismantling the sampling system

Sampling system

Peristaltic pumpfour-channel, 0-100 rpm
Sprayerconcentric, 400 μl/min
Spray chamberScott's two-way chamber, Peltier cooling from room temperature to -15 °C
Burner.low-consumption, solid, 2.4 mm injector, support for hot and cold plasma

Sample entry kits and consumables

Starter Kit PQMS AQUA

Standard sample entry kit for water samples

Starter Kit PQMS ORGANIC

Set for organic solvents and organic matrices

Starter Kit PQMS INERT Semicon

Inert corrosion-resistant PFA kit with platinum injector for semiconductor and high-purity tasks

Platinum interface cones

For petrochemical samples and aggressive matrices

Options and modules

Autosamplers

For high-performance automatic analysis of sample series

Aerosol breeding

Automatic dilution of aerosol samples for working with high-matric samples without manual dilution

Nitrox

Gas console for direct analysis of pure organic solvents

Clarity HPLC Software

Software module for interface with HPLC