ISP-MS spectrometer PlasmaQuant MS
Sustainable ISP-MS for high-matric samples: sensitivity of more than 500 thousand IMP / s per μg / l
PlasmaQuant MS is a compact ISP-MS for reliable multi-element analysis of complex and high-matrix samples, including biological fluids, food products and environmental samples.
Request a quotationThe device combines stable plasma with 90-degree reflective ion optics, which separates sensitive components from a straight line of photons and neutral particles. the iCRC's built-in collision reaction cell helps suppress spectral interference, and the open layout simplifies access to the sample input system and daily maintenance. The model is focused on stable work with high mineralization matrices.
The declared sensitivity for 115In is more than 500 thousand IMP / s per μg / l, the background at a mass of 5 AU is less than 0.5 IMP / s. The total platform of the series includes a four-channel peristaltic pump 0-100 RPM, a concentric sprayer with a flow rate of 400 μl / min, a two-way Scott chamber with Peltier cooling from room temperature to −15 ° C and a single-piece. low-consumption burner with an injector of 2.4 mm The manufacturer indicates the consumption of plasma-forming argon about 10 l / min and stability for high-matrix samples.
Ion optics and separation
- ReflexION’s 90-degree reflective ion optics with three-dimensional ion beam focus separates sensitive components from a direct beam of photons and neutrals
- HD-quadrupole with a full mass range of 3-260 AU provides high resolution and low abundance sensitivity
- The all-digital AD Detection does not require regular cross-calibration, as is the case with analog-to-digital detectors.
Elimination of interference and stability
- The built-in helium and hydrogen iCRC reaction cell provides rapid and efficient suppression of molecular interferences.
- BOOST technology supports sensitivity in gas mode without degrading detection limits
- Interface cones are replaced in less than a minute without disassembly of the sample input system, which reduces downtime
Sampling and software
- The open layout of the input system simplifies access to the sprayer, camera and burner during everyday maintenance.
- ASpect MS software automates calibration and resolution tests
- Modules for aerosol dilution, work with pure organic solvents (Nitrox) and pairing with HPLC are available
Sensitivity and specificity of the model
| Parameter | PlasmaQuant MS |
|---|---|
| Sensitivity (115In) | more than 500,000 IMP/s per μg/L |
| Background (mass 5 AU) | less than 0.5 mp/s |
| Argon flow (plasma-forming) | about 10 l/min |
| Purpose | stable characterization of high-matric samples |
Ion optics and detector
| Parameter | PlasmaQuant MS |
|---|---|
| Ion optics | ReflexION, 90° reflection, 3D focusing of the ion beam |
| quadrupole | HD-quadrupole 3 MHz, mass range 3-260 AU, min. fixation time 50 μs, scanning speed > 5000 AU / s |
| RF generator | solid-state, 27 MHz, plasma power 300-1600 W |
| Detector | All-digital AD Detection system |
Elimination of interference
| Parameter | PlasmaQuant MS |
|---|---|
| Reaction cell | iCRC (Integrated Collision Reaction Cell), He and H2 gases at the top of the skimmer cone |
| BOOST technology | maintains sensitivity in gas mode without loss of detection limits |
| Replacement of interface cones | less than 1 min without dismantling the sampling system |
Sampling system
| Parameter | PlasmaQuant MS |
|---|---|
| Peristaltic pump | four-channel, 0-100 rpm |
| Sprayer | concentric, 400 μl/min |
| Spray chamber | Scott's two-way chamber, Peltier cooling from room temperature to -15 °C |
| Burner. | low-consumption, solid, 2.4 mm injector, support for hot and cold plasma |